Philos DR-T Technical Manual 

19

 

 

AV Hysteresis 

Off

; low; medium; high 

AV Repetitive Hysteresis 

Off

; 1...(1)...6 

AV Scan Hysteresis 

Off

; 1...(1)...6 

AV safety delay (Dual chamber only) 

100 ms 

Sense Compensation 

Off

; -15...(15)...-120 ms 

Ventricular Blanking Time 

16; 24; 32; 40; 48; 56; 72 ms 

Magnet effect 

Automatic

; asynchronous; synchronous 

Asynchronous Magnet Effect: paces at 90 ppm. 

Automatic Magnet Effect; 10 cycles at 90 

ppm 

asynchronous; thereafter synchronous with the programmed 
basic rate  

Synchronous Magnet Effect; synchronous with programmed 
basic rate  

Pulse amplitude 

A

 0.1...(0.1)...

3.6

...(0.1)...4.8...(1.2)...8.4 V 

V

 0.1...(0.1)...

3.6

...(0.1)...4.8...(1.2)...8.4 V 

Ventricular Pulse Amplitude for Safe Program 

4.8V 

Pulse width 

A

 

0.1; 0.2; 0.3;

 0.4

; 0.5; 0.75; 1.0; 1.5 ms 

V

 

0.1; 0.2; 0.3; 

0.4

; 0.5; 0.75; 1.0; 1.5 ms 

Содержание Philos DR-T

Страница 1: ...Philos DR T DDDR Dual Chamber Pulse Generator with Home Monitoring Technical Manual...

Страница 2: ...dards lead pulse generator compatibility should be confirmed with the pulse generator and or lead manufacturer prior to the implantation of a pacing system IS 1 wherever stated in this manual refers t...

Страница 3: ...4 3 Patient Message 13 5 Description of Transmitted Data 15 5 1 The Monitoring Interval 15 5 2 Heart Rate 15 5 3 Atrial Rhythm 15 5 4 Ventricular Rhythm 15 5 5 AV Conduction 16 5 6 System Status 16 6...

Страница 4: ...ii Philos DR T Technical Manual CAUTION Federal U S A law restricts this device to sale by or on the order of a physician or properly licensed practitioner...

Страница 5: ...e physician optimize the therapy process as it may result in the patient being scheduled for additional clinical appointments between regular follow up visits if necessary The implant s Home Monitorin...

Страница 6: ...them during his or her occupational and leisure activities The patient device comes with a rechargeable battery that has an approximate operational time of 24 hours after a charge time of 5 hours It r...

Страница 7: ...lected during registration of the patient For more information on registering for Home Monitoring contact your BIOTRONIK sales representative The password protected BIOTRONIK Home Monitoring website c...

Страница 8: ...Netscape Navigator 4 72 MS Internet Explorer 5 5 or Netscape 7 Mozilla Acrobat Reader Version 4 Version 5 or higher Communication Channel Fax G3 or e mail Fax G3 e mail or mobile phone Additionally t...

Страница 9: ...aindications For the general indications and contraindications please refer to the Philos Family technical manual The indications and contraindications of the Philos DR T are identical to those of the...

Страница 10: ...6 Philos DR T Technical Manual...

Страница 11: ...Output Settings High output settings combined with extremely low lead impedance may reduce the life expectancy of the pulse generator to less than 1 year Programming of pulse amplitudes higher than 4...

Страница 12: ...essage function is activated Otherwise this function will not be available Not for Diagnosis The data transmitted by Home Monitoring are not suitable for diagnosis because not all information availabl...

Страница 13: ...Philos DR T Technical Manual 9 Consult the Philos manual for additional warnings and precautions associated with this device...

Страница 14: ...10 Philos DR T Technical Manual...

Страница 15: ...ediately Patient Message the patient initiates transmission of a message The attending physician decides whether or not the patient will be able to initiate the transmission of a message on their own...

Страница 16: ...hen ERI mode is reached this status is transmitted Further measurements and transmissions of Home Monitoring data are no longer possible NOTE The attending physician must notify the BIOTRONIK Service...

Страница 17: ...or other time when the patient is usually in his or her home The length of the time interval monitoring interval is not programmable it is preset to daily 4 3 Patient Message It is possible to trigger...

Страница 18: ...14 Philos DR T Technical Manual...

Страница 19: ...set at 24 hours For an event or patient message the monitoring interval would typically be less than 24 hours This occurs when these messages are sent after the programmed transmission time of the tr...

Страница 20: ...ntricular stimulation AsVp with dual chamber stimulation ApVp 5 6 System Status Atrial lead check Ventricular lead check Mean P Wave amplitude programmed sensitivity Mean R Wave amplitude programmed s...

Страница 21: ...A DDT V DDT DDI DDI T DVI DVT DOO VDD VDT VDI VVI VVT VOO AAI AAT AOO OFF The Home Monitoring function is available for the following pacing modes DDDR DDTR A DDTR V DDTR DDIR DDIR T VDDR VDTR VDIR DD...

Страница 22: ...on Rate Limitation2 3 4 190 220 ppm Dynamic AV Delay Dual chamber only low medium high individual fixed AV Delay Values Dual chamber only 15 50 75 100 120 10 200 225 250 300 ms Programmable in 5 range...

Страница 23: ...Automatic asynchronous synchronous Asynchronous Magnet Effect paces at 90 ppm Automatic Magnet Effect 10 cycles at 90 ppm asynchronous thereafter synchronous with the programmed basic rate Synchronous...

Страница 24: ...Automatic Mode Conversion Off On in modes DDD R DDT R A DDT R V and VDD R Mode Switch X out of Y Off On in modes DDD R DDT R A DDT R V and VDD R X 3 1 8 Z 3 1 8 Lead Polarity Pace A unipolar bipolar...

Страница 25: ...hreshold very low low mean high very high Rate increase 1 2 4 8 ppm s Maximum sensor rate 80 5 180 ppm Rate decrease 0 1 0 2 0 4 0 8 ppm s Automatic Sensor Gain Off On 6 3 2 Parameters at Replacement...

Страница 26: ...basic rate at 80 ppm Synchronized with basic rate reduced by 4 5 11 Asynchronous Asynchronous basic rate at 80 ppm Asynchronous with basic rate at 80 Synchronous Synchronized with basic rate reduced b...

Страница 27: ...battery pulse and lead data Two channel Real Time IEGM Transmission with markers Patient Data Memory Sensor Simulation Position Indicator for the programmer head 24 hour Trend Heart Rate Histogram Se...

Страница 28: ...can be configured with the following programming devices EPR 1000PLUS TMS 1000PLUS 6 5 Materials in Contact with Human Tissue Housing Titanium Connector receptacle Epoxy resin Sealing Plugs Silicone R...

Страница 29: ...ce uncoated 32 8 cm2 coated 7 23 cm2 Conducting Shape uncoated flattened ellipsoidal coated ellipsoidal BATTERY Power Source Li I Open Circuit Voltage 2 8 V Voltage at ERI 2 5 V Nominal Capacity6 1 3...

Страница 30: ...26 Philos DR T Technical Manual Volume 12 cm3 Dimensions 6 x 44 x 51 mm X Ray Identification VV...

Страница 31: ...Philos DR T Technical Manual 27 7 Order Information Product Name Order Number Philos DR T 331 440...

Страница 32: ...Distributed by BIOTRONIK Inc 6024 Jean Road Lake Oswego OR 97035 5369 800 547 0394 24 hour 503 635 9936 FAX Manufactured by BIOTRONIK GmbH Co KG Woermannkehre 1 D 12359 Berlin Germany M4064 C 7 04...

Отзывы: