31
Parameters
Pacing and Sensing
Lead configuration
IEGM recordings
Rates for statistics
Parameter
Range of values
Standard
SR
DR HF
Sensing polarity A
Unipolar; bipolar
Unipolar
x
x
x
Sensing polarity RV
Unipolar; bipolar
Unipolar
x
x
x
Sensing polarity LV
Unipolar; bipolar
Unipolar
x
Pacing polarity A
Unipolar; bipolar
Unipolar
x
x
x
Pacing polarity RV
Unipolar; bipolar
Unipolar
x
x
x
Pacing polarity LV
Device type HF:
LV1 tip -> LV2 ring
LV1 tip -> RV ring
LV2 ring -> LV1 tip
LV2 ring -> RV ring
LV1 tip -> housing
LV2 ring -> housing
LV1 tip –>
housing
x
Device type HF QP
LV1 tip -> LV2 ring
LV1 tip -> LV4 ring
LV1 tip -> RV ring
LV1 tip -> housing
LV2 ring -> LV1 tip
LV2 ring -> LV4 ring
LV2 ring -> RV ring
LV2 ring -> housing
LV3 ring -> LV2 ring
LV3 ring -> LV4 ring
LV3 ring -> RV ring
LV4 ring -> LV2 ring
LV4 ring -> RV ring
LV1 tip –>
LV2 ring
x
Parameter
Range of values
Standard
SR
DR HF
Number of recordings
(each max. 10 s)
4 series: 4
12
8 series: 20
—
x
x
x
High atrial rate (HAR) OFF; AT; mode switching
AT
x
x
x
High ventricular rate
(HVR)
OFF; ON
ON
x
x
x
8 series:
Patient triggering
(triggered by patient)
OFF; ON
OFF
x
x
x
Pre-trigger recording 0; 25; 50; 75; 100%
75%
x
x
x
IEGM signal
Filtered; Unfiltered
Filtered
x
x
x
Parameter
Range of values
Standard
SR
DR HF
HAR limit
100 ... (10) ... 250 bpm
200 bpm
x
x
HVR limit
150 ... (5) ... 200 bpm
180 bpm
x
x
x
HVR counter
4; 8; 12; 16 events
8 events
x
x
x
Start resting period
00:00 ... (1:00 AM) ...
23:00 hh:mm
2:00 hh:mm
x
x
x
Duration of resting
period
0.5 ... (0.5) ... 12 h
4 h
x
x
x
Enable lead check
OFF; ON
ON
x
x
x