29
Parameters
Timing
Refractory periods
Blanking periods
PMT protection
Parameter
Range of values
Standard
SR
DR HF
RV refractory period
200 ... (25) ... 500 ms
250 ms
x
x
x
Atrial refractory
period
AUTO
AUTO
x
x
Atrial refractory
period in the modes
AAI(R); AAT(R); DDT
300 ... (25) ... 775 ms
350 ms
x
x
LV refractory period
200 ms
200 ms
x
AUTO PVARP
OFF; ON
ON
x
x
PVARP
175 ... (25) ... 600 ms
225 ms
x
x
PVARP after PVC
PVARP + 150 ms
(max: 600 ms)
Automatically set
x
x
Parameter
Range of values
Standard
SR
DR HF
Far-field protection
after Vs
100 ... (10) ... 220 ms
100 ms
x
x
Far-field protection
after Vp
100 ... (10) ... 220 ms
150 ms
x
x
Ventricular blanking
period after Ap
30 ... (5) ... 70 ms
30 ms
x
x
Parameter
Range of values
Standard
SR
DR HF
PMT protection
OFF; ON
ON
x
x
VA criterion
250 ... (25) ... 500 ms
350 ms
x
x