BIOTRONIK BIOwand Скачать руководство пользователя страница 15

Getting Started

Establishing a Connection to a stimulator

13

To establish a Bluetooth connection to the BIOwand app for the first time, proceed as follows:
1. Switch on the device with the Bluetooth interface on which the BIOwand app is installed.
2. Ensure that BIOwand is ready for use.
3. On the device on which the BIOwand app is installed, open the menu to establish a Bluetooth

connection.

4. View all available Bluetooth devices.
5. Select the desired BIOwand.

The display name of the respective BIOwand is printed on its type plate.

6. In the menu, select the Connect menu item.

The Bluetooth status indicator (LED) on the BIOwand starts flashing blue.

7. Within a max. 30 s, press the Pair/Disconnect key of the BIOwand to establish the Bluetooth

connection.
The Bluetooth status indicator (LED) lights up blue continuously.

Disconnecting the BIOwand Bluetooth connection

To disconnect the Bluetooth connection, proceed as follows:
1. Press the Pair/Disconnect key of the BIOwand to disconnect the Bluetooth connection.

The Bluetooth status indicator (LED), which lights up blue continuously, turns off.

Establishing a Connection to a stimulator

To establish a connection to a stimulator, proceed as follows:
1. Ensure that there is a Bluetooth connection to the device on which the BIOwand app is installed.
2. Ensure that the BIOwand app is installed and running.
3. Place the programming head on the patient, at the point where the stimulator is implanted or the

external stimulator is attached.

4. Ensure the connection quality between the programming head and the stimulator using the

telemetry status indicator (ring LED) on the PGH, see Telemetry Status Indicator on the

Programming Head [Page 11].

5. Follow all further instructions and proceed as described in the documentation for the BIOwand

app.

Содержание BIOwand

Страница 1: ...BIOwand Device for wireless communication with the BIOTRONIK stimulator Technical Manual 452051 Revision C DRAFT 2021 02 26...

Страница 2: ...istributor BIOTRONIK Inc 6024 Jean Road Lake Oswego OR 97 035 5369 Tel 800 547 0394 24 hour Fax 503 635 9936 marketing biotronikusa com BIOTRONIK SE Co KG All rights reserved Specifications subject to...

Страница 3: ...tus Indicator LED 11 Power Adapter On Off Light Indicator LED 11 Bluetooth Status Indicator LED 11 Telemetry Status Indicator on the Programming Head 11 Setting up the Device 12 Connecting the Program...

Страница 4: ...cation are identical to those of the SCS system Required expertise The use of the device and this technical manual are intended for field representatives of BIOTRONIK who are familiar with the followi...

Страница 5: ...aulty operation of the device Only use accessories authorized by BIOTRONIK Precautions Risk of electromagnetic interference The use of this device close to or in direct contact with other devices shou...

Страница 6: ...ices and components Liquids Never connect a damp or wet device to the mains supply and never use damp or wet components Protect the device and its components from the accidental ingress of liquids and...

Страница 7: ...nt at the same time Power supply The device is operated via the provided power adapter The electrical outlet must fulfill the following conditions The mains installation at least fulfills the requirem...

Страница 8: ...formed according to IEC 60601 1 2 2014 Section Test Test level 7 1 EN 55011 CISPR 11 Conducted interference emissions Group 1 Class B EN 55011 CISPR 11 Radiated emission 7 2 1 IEC 61000 3 2 Harmonic d...

Страница 9: ...promise the function of the device in certain cases This kind of device malfunction should be considered as a possible cause if the following is observed The device switches on by itself The device di...

Страница 10: ...ollowing regulations are applicable to the device and its components Disconnect the power plug from the outlet before cleaning and or disinfecting Use lint free soft cloths Clean the housing with a da...

Страница 11: ...Owand PGH 3 Power adapter FW8000M 12 with on off light indicator and power plug 4 Plug for connecting the power adapter to the device BIOwand Device 5 Programming head connection BIOwand PGH 6 Pair di...

Страница 12: ...manual Position for the guide cam of the Redel plug Device contains materials that must be correctly disposed of in accordance with environ mental protection regulations The European Directive 2012 19...

Страница 13: ...icator LED The status indicator for pairing disconnecting the Bluetooth connection shows the following device statuses LED behavior and color Device status LED does not indicate any behavior There is...

Страница 14: ...he power adapter FW8000M 12 and to switch on the device proceed as follows 1 Select the appropriate power plug adapter for your region for the enclosed power adapter 2 Insert the required power plug a...

Страница 15: ...ghts up blue continuously Disconnecting the BIOwand Bluetooth connection To disconnect the Bluetooth connection proceed as follows 1 Press the Pair Disconnect key of the BIOwand to disconnect the Blue...

Страница 16: ...g head BIOwand PGH is connected 1 Check whether the programming head BIOwand PGH has been correctly connected 2 Connect a programming head BIOwand PGH An internal error has occurred 1 To switch off th...

Страница 17: ...essure 700 1060 hPa Operation at altitudes Up to 3 000 m Power supply to the device Operation with power adapter FW8000M 12 Physical properties Category Design Dimensions W x D x H 120 x 150 x 42 8 mm...

Страница 18: ...complete system device power adapter and programming head Country related information International radio certification Telemetry information for Australia This device is in compliance with the Austr...

Страница 19: ...ocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult...

Страница 20: ...ins materials that must be correctly disposed of in accordance with environ mental protection regulations The European Directive 2012 19 EU on waste electrical and electronic equipment WEEE 2 applies...

Отзывы: