To set up the BIOwand, proceed as follows:
1. Place the device and its components on a flat, dry surface.
2. Make sure that the device and its components are protected against liquids and moisture.
3. Make sure that the device cannot shift, even with the cable connected, and that the patient can only

come into contact with the programming head. Do not touch any plug connections such as Redel

connectors and the patient at the same time.

Connecting the Programming Head (BIOwand PGH)

The port for the programming head is located on the right rear side of the device. To connect the

BIOwand PGH, proceed as follows:
1. Insert the Redel plug of the programming head cable into the programming head connector on the

device. Pay attention to the correct position of the guide cam of the Redel plug. It clicks into place.

Connect the programming head so that the cable is resting with some slack in the cable and there is no

danger of tripping.

Connecting the Power Adapter and Switching on the Device

The device switches on automatically as soon as the provided power adapter (FW8000M/12) is

connected. The package contents include various country-specific adapters for the power plug. The

port for the power adapter is located on the left rear side of the device.
To 

connect the power adapter

 (FW8000M/12) and to 

switch on

 the device, proceed as follows:

1. Select the appropriate power plug adapter for your region for the enclosed power adapter.
2. Insert the required power plug adapter into the power adapter. It clicks into place.
3. Connect the power adapter to the outlet.
4. Insert the plug of the power adapter cable into the power supply port of the device.

The device performs a self-test and the BIOwand status indicator (LED) flashes green. After

successful self-test, the BIOwand status indicator (LED) lights up green continuously. If the self-

test is not successful, the BIOwand status indicator (LED) lights up orange continuously, see Error

Resolution [Page 14].

Connect the power adapter so that the cable is laid without tensile stress and there is no danger of

tripping. Ensure that the power plug is freely accessible at all times.
To 

switch off

 the device, proceed as follows:

1. Disconnect the power plug of the power adapter from the wall outlet.

The BIOwand status indicator (LED) of the device turns off.

Establishing a Connection to the BIOwand App

Note

The display name of the respective BIOwand is printed on its type plate on the underside of the

device. It consists of the product name "BIOwand" and the last three digits of the serial number

(e.g., BIOwand678).

The Bluetooth connection must be established manually the first time the BIOwand is used. After the

Bluetooth connection has been established, BIOwand automatically connects to the device on which the

BIOwand app is installed as soon as BIOwand is ready for use.

Содержание BIOwand

Страница 1: ...BIOwand Device for wireless communication with the BIOTRONIK stimulator Technical Manual 452051 Revision C DRAFT 2021 02 26...

Страница 2: ...istributor BIOTRONIK Inc 6024 Jean Road Lake Oswego OR 97 035 5369 Tel 800 547 0394 24 hour Fax 503 635 9936 marketing biotronikusa com BIOTRONIK SE Co KG All rights reserved Specifications subject to...

Страница 3: ...tus Indicator LED 11 Power Adapter On Off Light Indicator LED 11 Bluetooth Status Indicator LED 11 Telemetry Status Indicator on the Programming Head 11 Setting up the Device 12 Connecting the Program...

Страница 4: ...cation are identical to those of the SCS system Required expertise The use of the device and this technical manual are intended for field representatives of BIOTRONIK who are familiar with the followi...

Страница 5: ...aulty operation of the device Only use accessories authorized by BIOTRONIK Precautions Risk of electromagnetic interference The use of this device close to or in direct contact with other devices shou...

Страница 6: ...ices and components Liquids Never connect a damp or wet device to the mains supply and never use damp or wet components Protect the device and its components from the accidental ingress of liquids and...

Страница 7: ...nt at the same time Power supply The device is operated via the provided power adapter The electrical outlet must fulfill the following conditions The mains installation at least fulfills the requirem...

Страница 8: ...formed according to IEC 60601 1 2 2014 Section Test Test level 7 1 EN 55011 CISPR 11 Conducted interference emissions Group 1 Class B EN 55011 CISPR 11 Radiated emission 7 2 1 IEC 61000 3 2 Harmonic d...

Страница 9: ...promise the function of the device in certain cases This kind of device malfunction should be considered as a possible cause if the following is observed The device switches on by itself The device di...

Страница 10: ...ollowing regulations are applicable to the device and its components Disconnect the power plug from the outlet before cleaning and or disinfecting Use lint free soft cloths Clean the housing with a da...

Страница 11: ...Owand PGH 3 Power adapter FW8000M 12 with on off light indicator and power plug 4 Plug for connecting the power adapter to the device BIOwand Device 5 Programming head connection BIOwand PGH 6 Pair di...

Страница 12: ...manual Position for the guide cam of the Redel plug Device contains materials that must be correctly disposed of in accordance with environ mental protection regulations The European Directive 2012 19...

Страница 13: ...icator LED The status indicator for pairing disconnecting the Bluetooth connection shows the following device statuses LED behavior and color Device status LED does not indicate any behavior There is...

Страница 14: ...he power adapter FW8000M 12 and to switch on the device proceed as follows 1 Select the appropriate power plug adapter for your region for the enclosed power adapter 2 Insert the required power plug a...

Страница 15: ...ghts up blue continuously Disconnecting the BIOwand Bluetooth connection To disconnect the Bluetooth connection proceed as follows 1 Press the Pair Disconnect key of the BIOwand to disconnect the Blue...

Страница 16: ...g head BIOwand PGH is connected 1 Check whether the programming head BIOwand PGH has been correctly connected 2 Connect a programming head BIOwand PGH An internal error has occurred 1 To switch off th...

Страница 17: ...essure 700 1060 hPa Operation at altitudes Up to 3 000 m Power supply to the device Operation with power adapter FW8000M 12 Physical properties Category Design Dimensions W x D x H 120 x 150 x 42 8 mm...

Страница 18: ...complete system device power adapter and programming head Country related information International radio certification Telemetry information for Australia This device is in compliance with the Austr...

Страница 19: ...ocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult...

Страница 20: ...ins materials that must be correctly disposed of in accordance with environ mental protection regulations The European Directive 2012 19 EU on waste electrical and electronic equipment WEEE 2 applies...

Отзывы: