104
|
Chapter 6: Instrument Qualification
Repeatability Results
Wells C1 E2 G3 H6 F5 D4
Read 1 0.134 0.566 1.037 1.557 1.866 2.385
Min Limit 0.128 0.555 1.022 1.536 1.842 2.308
Max Limit 0.140 0.577 1.052 1.578 1.890 2.462
Read 2 0.135 0.566 1.037 1.558 1.867 2.385
Result PASS PASS PASS PASS PASS PASS
Reviewed/Approved By: _______________________________________
Date: ________________
For Technical Support
In the U.S.: In Europe:
BioTek Instruments, Inc. BioTek Instruments GmbH
Tel: 800 242 4685 Tel: 49 (0) 7136-9680
Fax: 802 655 3399 Fax: 49 (0) 7136-968-111
All Others:
Tel: 802 655 4040
Fax: 802 655 3399
email: [email protected]
Product support center: www.biotek.com/service
Figure 2: Sample output for the Absorbance Plate Test
Results and Troubleshooting Tips
The Absorbance Test Plate Report contains results for the following:
•
Peak Absorbance:
When the test is performed, the C6 filter is scanned at the
test range(s) defined by the user in the Absorbance Test Plate dialog. To verify
wavelength accuracy, the wavelength of the maximum absorbance is compared
with the peak wavelength value entered in the software, which comes from the
data sheet supplied with the Test Plate. The accuracy of the wavelength should
be ± 3 nm (± 2 nm instrument, ± 1 nm filter allowance).
If the test fails:
Check the C6 filter to make sure it is clean. If needed, clean it with lens
paper.
Important!
Do not remove the filter from the Test Plate, and do
not use alcohol or other cleaning agents.
Make sure the information entered into Gen5 matches the information on
the Test Plate’s data sheet.
Make sure the Test Plate is within its calibration certification period. The
calibration sticker is affixed directly to the plate. If it is out of date, contact
BioTek to schedule a recertification.
Check the microplate carrier to ensure it is clear of debris.
BioTek Instruments, Inc.
Содержание Synergy HTX
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Страница 2: ......
Страница 12: ...viii Preface Revision History Rev Date Changes A 7 2014 Initial release BioTek Instruments Inc ...
Страница 22: ...xviii Preface BioTek Instruments Inc ...
Страница 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 31: ...1 Unpack and Inspect the Reader 9 Figure 1 Unpacking the reader accessories box Synergy HTX Operator s Manual ...
Страница 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Страница 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Страница 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Страница 166: ......
Страница 174: ...150 Appendix A Specifications BioTek Instruments Inc ...