Figure 4-5: Removing the screw located between the EM1 and
EM2 positions
4. Carefully lift the top off the filter cube.
Figure 4-6: Filter cube with the top removed (left), exposing the
mirrors (right)
Caution:
Do not touch the mirrors with your bare fingers. Wear gloves
to reduce the risk of damaging the mirrors or polarizing filters. If you
accidentally touch a mirror or polarizing filter with your bare fingers,
see
starting on page
for cleaning
instructions.
Caution:
When removing or replacing a filter or C-clip filter retainer,
do
not use a sharp tool
. Use several layers of lens paper and your finger or
a cotton swab to remove and replace filters and clips. Using a sharp
instrument, such as a flat screwdriver, will scratch the filter surface and
make it unusable.
50 | Chapter 4: Filters and Mirrors
BioTek Instruments, Inc.