Synergy H1 Abs Test 2.prt
Parameter
Setting
Plate Type
96 WELL PLATE
Shake Step
Linear, 4 minutes, default frequency
Two Read Steps
Kinetic loop (one per Read step)
Set a Run Time/Interval combination to read the
plate five times with minimal delay
Detection Method
Absorbance
Read Type
Endpoint
Optics Type
Monochromators
Step labels
First Read step: "Normal"
Second Read step: "Turnaround"
Read wells
Full plate
Wavelengths
2 (450 nm, 630 nm)
Read Speed
Normal
Delay after plate movement
100 msec
Data Reduction
Define two Delta OD transformations (450-630
nm), one per Read data set
Synergy H1 Abs Test 3.prt
Parameter
Setting
Plate Type
96 WELL PLATE
Shake Step
Linear, 30 seconds, default frequency
Kinetic loop
Set a Run Time/Interval combination to read the
plate five times with minimal delay
Detection Method
Absorbance
Read Type
Endpoint
Optics Type
Monochromators
Read wells
A1..H6
Wavelength
340 nm
Read Speed
Normal
Delay after plate movement
100 msec
Synergy H1 Operator's Manual
Absorbance Testing | 79
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...