134 | Chapter 3: Operation
Handling Special Plates and Mini-tubes
Note:
Also see the
.
l
Peri-pump Special Plate Handling below
l
1536F - 1536-well Flanged Plates
Important:
Crashes can occur! Remove unused manifolds when dispensing to 153F
plates.
To prevent an unused Peri-pump cassette or Syringe manifold from colliding with
the plate flange during dispensing:
l
use
both
Syringe manifolds simultaneously, and,
l
unload the Peri-pump cassette when it is not being
used;
l
alternatively, remove the Syringe manifolds while
using the Peri-pump;
l
limit protocols to one dispenser only, either the
Syringe or Peri-pump, i.e. run multiple protocols on a
plate when the assay requires using more than one
device.
To achieve the best dispense performance when processing 1536-well flanged
plates, the MultiFlo FX's default protocol settings specify a low dispense height to
position the dispense tubes as close as possible to the wells. When the dispense
arm holds multiple dispense manifolds, the manifolds not addressing the wells can
collide with the plate flanges during a dispense.
Peri-pump Special Plate Handling
Note:
Applies to non-RAD cassettes only, i.e. standard 8-channel cassettes.
The
Peri-pump
supports several plate types, but vessels with fewer than 8 rows
require special handling. Some adjustment of or consideration of how the 8-channel
dispense head will address the plate is needed.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......