72 | Instrument Qualification
BioTek Instruments, Inc.
System Self-Check and Checksum Test
System Self-Check
Each time it is turned on, the MicroFlo Select automatically runs the System Self-Check
to verify system components. The self-check verifies movement of the dispense arm
and plate carrier.
Using the Keypad
Select
UTIL> SLFCHK
Using the LHC
Select
Tools> Instrument Utilities> Self Check
When the test completes successfully a “passed” message is displayed. If the test fails,
the dispenser will “chirp” and display an error code.
Checksum Test
Using the Keypad
Select
UTIL> CHKSUM
Select
UI
, record the version code displayed and press
Enter
to return to the
Checksum options. Repeat for
MC
.
Using the LHC
Select
Tools> Instrument Utilities> Version
(click the Version link to refresh the
checksum information.
Record the version and checksum information.