Clean Exposed Surfaces | 39
800 TS
Warning!
Mucous membranes are considered prime entry routes for infectious
agents. Wear eye protection and a surgical mask when there is a possibility of
aerosol contamination. Intact skin is generally considered an effective barrier
against infectious organisms; however, small abrasions and cuts may not always be
visible. Wear protective gloves when handling contaminated instruments.
Clean Exposed Surfaces
Exposed surfaces may be cleaned (not decontaminated) with a cloth moistened (not soaked)
with water or water and a mild detergent. You’ll need:
•
Mild detergent
•
Deionized or distilled water
•
Clean, lint-free cotton cloths
1.
Turn off and unplug the instrument.
2.
Wet a clean cotton cloth with water, or with water and mild detergent, then
thoroughly wring out the cloth so that liquid does not drip from it.
3.
Wipe the plate carrier, the inside of the plate carrier door (if equipped), and all exposed
surfaces of the instrument.
4.
If detergent was used, wipe all surfaces with a cloth moistened with water.
5.
Use a clean, dry, lint-free cloth to dry all wet surfaces.
Clean the Touchscreen
Materials
Use the following products to safely clean the touchscreen:
•
Deionized or distilled water
•
Dish soap or other mild cleaner
•
Lintfree disposable towels
This procedure is for the housing of the 800 TS instrument. See "Clean the Touch
Screen" on page 39 for the cleaning procedure for the touchscreen.
Important!
Never spray solutions directly on the touchscreen.
Содержание 800 TS Series
Страница 1: ...800 TS Absorbance Reader Instructions for Use BioTek Instruments Inc 2017 PN 1561011 Rev A ...
Страница 12: ...xvi Safety Symbols BioTek Instruments Inc ...
Страница 13: ...Installation ...
Страница 28: ...BioTek Instruments Inc 24 Repackaging and Shipping Instructions ...
Страница 29: ...Getting Started ...
Страница 41: ...Maintenance ...
Страница 48: ...BioTek Instruments Inc 44 Filter Storage and Handling ...
Страница 49: ...Instrument Testing ...
Страница 55: ...Specifications ...