TurboVap

®

User Manual

Содержание TurboVap

Страница 1: ...TurboVap User Manual...

Страница 2: ...Adjust the Nozzles 9 Set the Water Bath Temperature 10 Evaporate 10 Set Up and Start an Evaporation Run 11 Monitor an Evaporation Run 11 Edit an Evaporation Run in Progress 11 End of an Evaporation R...

Страница 3: ...the sample to assure that darkening or brightening of samples does not stop the evaporation prematurely Once the sample solution passes below the end point sensor the system stops the evaporation in...

Страница 4: ...d point sensors can only be used at bath temperatures up to 60 C The evaporation is automatically paused the gas is turned off when the lid is opened and resumed when the lid is closed The water bath...

Страница 5: ...s heating the water bath to the set starting bath temperature when started at scheduled wake up time or the last used water bath temperature when started manually Audible Alarm and Lighting To provide...

Страница 6: ...ntrate your samples by performing a test run using a tube with the approximate amount of solvent you want to evaporate Start with the gas flow set to the lowest setting and then slowly increase the ga...

Страница 7: ...segments Evaporate using end point detection see End Point and End Point Time above Evaporate using a combination of a gradient and end point detection Evaporate for a set period of time Notes The rac...

Страница 8: ...in the settings view see Figure 10 or automatically by scheduling a sleep time in the sleep mode view see Figure 11 The system will also enter sleep mode if the lid is left open for more than 90 minut...

Страница 9: ...e measured temperature 5 Repeat steps 2 through 4 with the second water bath temperature 6 When you are done press More in the settings view and then Calibrate Temp 7 Enter the values of set and measu...

Страница 10: ...e DRAIN port B Add deionized water until the water s surface is as high as the initial liquid level will be in the tubes but not higher than just below the blue water bath top cover Prepare the System...

Страница 11: ...e engraved adjustment lines on the manifold see Figure 16 Figure 16 The engraved adjustment lines on the manifold Set the Water Bath Temperature Warning The temperature of the water bath and the drain...

Страница 12: ...nt of time the evaporation continues in a sample tube after its liquid level passes below the sensor when in the end point time mode Gas Flow The internal gas flow directed at each of the sample tubes...

Страница 13: ...Finalizing 12 44 Wed 23 Nov 2016 Time Settings 54 min Time 31 min Time Elapsed Pause Stop 1 2 L min 75 C Edit Mode Gas Flow Bath Temperature Nozzle Setup 54 min L min 1 2 C 75 Status 54 00 min Total T...

Страница 14: ...where you select the method you want to run or modify a method and then run it Evaporation Parameters The following evaporation parameters are available in the edit method view see Figure 24 Gradient...

Страница 15: ...a week 1 To avoid the risk of burning yourself on the water or the drain connector lower the water bath temperature to below 55 C 2 If a rack is present in the water bath carefully lift it out of the...

Страница 16: ...r pops out see Figure 31 3 Clean the new fuse P N 415578SP using a cloth lightly dampened with ethanol and wipe it dry with a dry cloth Note Do not touch the metal surfaces after the fuse has been cle...

Страница 17: ...sample residues Avoid contact 1 Turn off the system and disconnect the power cord 2 Turn off the gas supply 3 If a rack is present in the water bath carefully lift it out of the bath clean it using a...

Страница 18: ...s for all rack positions that contain sample tubes are enabled in the software If there is no or slow evaporation in a single tube When using a system with TurboVap II configuration ensure that the no...

Страница 19: ...ption Qty C48985 10 x 75 mm Test Tube 1000 C44651 12 x 75 mm Test Tube 1000 C40707 13 x 100 mm Test Tube 1000 413282 16 x 75 mm Test Tube 1000 C40708 16 x 100 mm Test Tube 1000 C45273 16 x 125 mm Test...

Страница 20: ...654 4917 Order Tel 1 704 654 4900 Order Fax 1 434 296 8217 ordermailbox biotage com Support Tel 1 800 446 4752 Outside US 1 704 654 4900 us 1 pointsupport biotage com JAPAN Tel 81 3 5627 3123 Fax 81...