Quick Start Guide to TurboVap

®

 II | Page 2

Part Number: 415878-B

© 2020 Biotage. All rights reserved. No material may be reproduced or published without the written permission of Biotage.  

Information in this document is subject to change without notice and does not represent any commitment from Biotage. E&OE.  

A list of all trademarks owned by Biotage AB is available at www.biotage.com/legal. Other product and company names mentioned 

herein may be trademarks or registered trademarks and/or service marks of their respective owners, and are used only for 

explanation and to the owners’ benefit, without intent to infringe.

Brightness (11)

Display shows the brightness of the touch screen. Brightness 
is controlled by pressing either 

+

 or 

.  

Manifold Setup (12)

Display shows the selected/installed manifold. For more 
information, see “Manifolds” in the TurboVap

®

 User Manual.

Sleep Mode (13)

Display shows the sleep mode settings. For more information, 
see “Sleep Mode and Automatic Wake-Up” in the TurboVap

®

 

User Manual.

Temp Unit (14)

Display shows the temperature unit used by the system.  
Press 

Change

 to switch between °C or °F.

Time and Date (15) 

Display shows the current time and date. Time and date are 
set by pressing the appropriate blue buttons.

Lid Sleep (16)

With lid sleep turned ON, the system automatically enters 
sleep mode if the lid is left open for more than 90 minutes.

Cont End-Point (17)

Continuous end-point evaporation turned ON allows for 
replacing a finished tube and restarting end-point detection 
for that rack position without restarting the run.

Bath Light (18)

The bath light can be turned ON or OFF.

More (19)

Display shows system and software information, and gives 
access to calibration features. For more information, see the 
TurboVap

®

 User Manual.

Daily Operation

Follow the steps below to operate the system. For detailed 
instructions, see the TurboVap

®

 User Manual.

Prepare the System

The system should always be turned on when a sample tube is 
inserted so that the exhaust fan can remove solvent vapors.

1.  Check the incoming gas supply for sufficient reserve.
2.  Ensure that the water in the bath is at an appropriate level.
3.  Check/set the mode of operation.
4.  Check/set the water bath temperature. Red indicator in the 

Bath Temp

 field indicates bath heating. When the target 

water bath temperature has been reached, the user will be 
notified by a dialog on the touch screen.

5.  Check/set the gas flow.
6.  Open the lid and either load the sample tubes into the rack 

when positioned inside the water bath or gently lower the 
rack with the sample tubes into the water bath. The latter is 
not possible if using a rack with end-point sensors.

Note: 

If using a rack with end-point sensors, ensure the 

sensor connection box is connected to the 

AUX

 port on the 

right side of the system. 

7.  Close the lid.

Start the Evaporation Process

1.  Set the evaporation time if necessary.
2.  Enable the nozzles for the rack positions that contain sample 

tubes by pressing the 

Edit

 button in the 

Nozzle Setup 

field. 

3.  Press the 

Start

 button.

Stop/Pause the Evaporation Process

To pause the evaporation (stop and continue) for all tubes, 
press 

Pause

 or open the lid. To continue operation, press 

Resume

 or close the lid.

To stop the evaporation, press the 

Stop

 button. To stop the 

evaporation process for an individual sample tube, press 

Edit 

in 

the 

Nozzle Setup 

field and disable the nozzle for that tube.

Remove the Sample

Quick removal of a completed tube is necessary since highly 
volatile compounds can be lost if allowed to sit for an extended 
period.

Shut Down

When the system is no longer in use, leave the lid open to keep 
moisture from accumulating on the inside cover. Turn off the 
power (unless using sleep mode) and the gas supply (if desired).