Quick Start Guide to TurboVap

®

 II | Page 1

Evaporate

 Settings

Evaporating

15:51 Wed 22 Jan 2020

Edit

Evaporate

7:35 min

Done

6:53 min

Done

4:01 min

 Pause

 Stop 

End-Point + Time

 Mode

 Nozzle Setup

1.2 

L/min

L/min

1.2

15 min

 +Time

 Status

 Gas Flow

55 

°

C

°

C

55

 Bath Temp

45:55

Elapsed (min:sec)

1

2

6

9

3

4

5

8

7

For detailed instructions, see the TurboVap

®

 User Manual.

Mode (1)

Display shows the current operation mode. Mode is controlled 
by pressing the blue button. There are five evaporation modes 
available on TurboVap II:

 

»

Manual: 

Evaporate until the user presses the 

Stop

 button.

 

»

Time:

 Evaporate for a set period of time.

 

»

End-Point:

 Evaporate until the end-point sensors 

detect a liquid level of 0.2 or 0.7 mL depending on the 
glassware. Each sample tube is controlled individually.

 

»

End-Point + Time: 

Evaporate for a set period of time after the 

sensor detects a liquid level of 0.2 or 0.7 mL depending on 
the glassware. Each sample tube is controlled individually.

 

»

Method: 

There are several options. For more information, 

see “Define Methods” in the TurboVap

®

 User Manual.

Note that the two end-point detection modes are only enabled 
when a TurboVap II end-point rack has been connected to the 
system’s 

AUX

 port.

Time (2)

Display shows the set evaporation time (when in Time mode) 
or the amount of time the evaporation will continue in a sample 
tube after its liquid level passes below the sensor (when in 
End-Point + Time mode). (Not shown in other modes.)

Gas Flow (3)

Display shows the current and set gas flow in L/min. Gas flow is 
controlled by pressing the blue button.

Quick Start Guide to TurboVap

®

 II 

 

Bath Temperature (4)

Display shows the current and set water bath temperature in 
°C or °F. Bath temperature is controlled by pressing the blue 
button. Racks with end-point sensors can only be used at bath 
temperatures up to 60 °C.

Nozzle Setup (5)

Display shows which nozzles are currently enabled. Nozzles 
are selected ON or OFF by pressing the blue button. When in 
End-Point or End-Point + Time mode, selection can only be done 
when the system is not processing. 

Status (6)

Display shows the remaining and/or elapsed time of the current 
evaporation. When in End-Point or End-Point + Time mode, 
individual tube status is shown; OFF, Idle, Evaporate, or Done. 
When in End-Point + Time mode, remaining time is also shown.

Start/Pause (7)

Press to start or pause the evaporation in progress.

Stop (8)

Press to stop the evaporation in progress.

Settings (9)

Press to define system settings. Displays an alternative screen.

Evaporate

 Settings

System Ready

16:48 Wed 22 Jan 2020

 Manifold Setup

6 Positions

 Sound Level

Medium

    Sleep Mode

Edit

 Temp Unit

°

C

 Brightness

-

+

Change

 Time & Date

16

48

:

2020

Jan

22

Bath Light

OFF

ON

More »

Lid Sleep

OFF

ON

Cont End-Point

OFF

ON

10

11

14

15

18

12

13

19

17

16

Sound Level (10)

Display shows the current sound level. Sound level is controlled 
by pressing the blue button. The sound level can be set to Low, 
Medium, High, or OFF.