Error Information
Error Messages
Evaporator Temperature Limit Exceeded
Error ID: 19
Evaporator temperature exceeded threshold after stabilizing.
LED Light Source Error
Error ID: 20
The source has failed.
Evaporator Gas Flow Rate Limit Exceeded
Error ID: 21
Evaporator gas flow rate setting not met.
Probable cause
Suggested actions
1
Defective thermocouple
Please contact your Biotage service
representative
2
Defective evaporator heater
assembly
Please contact your Biotage service
representative
3
Defective main board
Please contact your Biotage service
representative
Probable cause
Suggested actions
1
LED light source not connected to
the main board
Please contact your Biotage service
representative
2
Defective LED light source
Please contact your Biotage service
representative
Probable cause
Suggested actions
1
Insufficient gas inlet pressure
Ensure that the gas inlet pressure is
above 60 psi
(4.1 bar)
2
Defective mass flow controller
Please contact your Biotage service
representative
Содержание ELSD-A120
Страница 1: ...Biotage ELSD A120 User Manual...
Страница 12: ...12 Biotage ELSD A120 User Manual 1 Introduction to the Detector Detector Overview...
Страница 20: ...20 Biotage ELSD A120 User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 44: ...44 Biotage ELSD A120 User Manual 5 Optimizing Performance Solvent Recommendations...
Страница 60: ...60 Biotage ELSD A120 User Manual 7 Error Information Error Messages...
Страница 74: ...74 Biotage ELSD A120 User Manual 8 Maintenance Replacing the Detector Firmware...
Страница 85: ......
Страница 86: ......