biosystems EX CHEK Скачать руководство пользователя страница 23

21

(9) Make sure the correct cylinder of gas is

attached before attempting to adjust the
span!  If the concentration of gas
reaching the sensor is too low to allow
the instrument to be adjusted, or if the
wrong type of gas is applied to the
sensor being adjusted, the alarms will be
activated and a screen will be displayed
indicating that the instrument “CANT /
SPAN”.  Pressing the MODE button
clears the “CANT SPAN” message and
returns the instrument to normal
operation.

C A N T

↑↓

S P A N

In the event that a “CANT SPAN”
message is displayed, verify the type and
concentration of the gas being used is
correct for the sensor being calibrated,
and replace the cylinder if necessary.
Verify that the flow rate of the regulator is
1.0 lpm.  If the "CANT SPAN" message is
repeated it may indicate that the sensor
requires replacement.

(10)

When span calibration has been

completed, press and hold the "Cal"
button down until the LCD screen
indicates that calibration is complete by
showing the message “DONE”.  The Ex
Chek will then turn itself off.

D O N E

The “Cal” button must be held down until
the screen indicates that span calibration
has been successfully completed.

If the button is released before this message
is displayed, span values will not be
updated, and remain unchanged from the
last time a span calibration was successfully
completed.
Pressing the MODE button at any time
cancels the manual span calibration
procedure.  A screen will announce “NO

SAVE” after which the instrument will turn
itself off.

N O

↑↓

S A V E

  A sensor that cannot be

calibrated or is found to be out of
tolerance should be replaced
immediately.  An instrument that fails
calibration may not be used until testing
with known concentration test gas
determines that accuracy has been
restored, and the instrument

 is once again

fit for use.

Pressing the MODE button at any time
during the span adjustment cancels the
procedure, and causes the calibration
values in memory to remain unchanged from
the last time a span calibration adjustment
was made.

Содержание EX CHEK

Страница 1: ...0 Reference Manual Ex Chek NiMH Rechargeable Version Combustible Gas Detector 651 South Main Street Middletown CT 06457 800 711 6776 860 344 1079 Fax 860 344 1068 Part number 13 123 Version 2 2...

Страница 2: ...1...

Страница 3: ...IS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED AVERTISSEMENT LIRE ATTENTIVEMENT LES INSTRUCTIONS AVANT DE METTRE EN MARCHE Ex Chek Reference Manual NiMH Version Biosystems Part Number 13 12...

Страница 4: ...t 9 1 9 3 Weather cover 10 1 10 Ex Chek design components 10 1 11 Ex Chek accessories 10 Chapter 2 Basic operation 11 2 1 Overview of Ex Chek operation 11 2 1 1 Operational warnings and cautions 11 2...

Страница 5: ...Chek Advanced Functions 24 5 1 Advanced features overview 24 5 2 Re setting the calibration gas concentration 24 5 3 Setting the security beep interval 24 Chapter 6 Trouble Shooting and Repair 25 6 1...

Страница 6: ...C T4 for instruments powered using DEMKO approved alkaline batteries The Ex Chek is also CE Marked Conformit Europ enne The UL CSA and DEMKO classifications marks indicate compliance with United State...

Страница 7: ...stored and the instrument is once again fit for use 6 Do not reset the calibration gas concentration setpoints in the Ex Chek unless the concentrations of your calibration gas differ from the concentr...

Страница 8: ...6...

Страница 9: ...h a range of zero to one hundred percent explosive The Ex Chek combustible gas sensor is non specific and responds to all combustible gases and vapors Combustible sensors contain two coils of fine wir...

Страница 10: ...n off MODE button once at any time during normal operation causes the backlight to be activated for 10 seconds In the event of a gas alarm the backlight turns on automatically and remains on until the...

Страница 11: ...ght and audible alarm are automatically evaluated and the battery is monitored continuously for proper voltage The Ex Chek offers the following additional options that allow the user to customize the...

Страница 12: ...as control most other functions 6 Protective cap An o ring sealed cap and moisture barrier filter protect the sensor from damage 7 Sensor The top mounted sensor may be easily accessed for replacement...

Страница 13: ...y can lead to inaccurate and potentially dangerous readings 5 A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument that fails calibration ma...

Страница 14: ...appear indicating the current peak alarm set point If the Ex Chek is operating in the Technician mode it will display the following factory installed default alarm set point P E A K A L R M 1 0 This...

Страница 15: ...ve Pressing the MODE button once displays the peak reading of the gas being measured since the instrument was switched on E X 2 0 Pressing the MODE button again returns the display to the first screen...

Страница 16: ...concentration test gas before the instrument is put back into service CAUTION To achieve optimal charge and ensure long battery life of the NIMH battery make sure that charging takes place in an area...

Страница 17: ...l stay in alarm mode until the unit is shut off and an operational sensor is installed Chapter 3 Calibration The Ex Chek detector has been designed for easy calibration The on off MODE button is used...

Страница 18: ...nsitivity is maintained over the life of the sensor 3 1 1 1 Effects of high concentrations of combustible gas on the combustible sensor The accuracy of combustible sensors may also be affected by expo...

Страница 19: ...rument using the procedures discussed in Section 3 4 1 Slip on the calibration adapter as discussed in Section 3 4 2 and flow calibration gas to the sensor Wait for the readings to stabilize Forty fiv...

Страница 20: ...l show the message WAIT ZERO while the adjustment is being completed W A I T Z E R O 6 After successful completion of the zero calibration the display will prompt you with the message SPAN MODE to con...

Страница 21: ...s the concentration shown on the Span Adjust screen as the sensor is adjusted If the concentration does not match it will be necessary to assign a new calibration gas concentration as discussed in Cha...

Страница 22: ...hat the operating mode is changed the LCD screen will briefly indicate the current operating mode 5 Turn the instrument off by pressing the MODE button for three full seconds 6 With the unit turned of...

Страница 23: ...the LCD screen indicates that calibration is complete by showing the message DONE The Ex Chek will then turn itself off D O N E The Cal button must be held down until the screen indicates that span c...

Страница 24: ...e Ex Chek are very conservative in order to maximize worker safety To enter the alarm adjust mode it is necessary to do the following 1 Turn the instrument on and make sure it is in Technician Mode Se...

Страница 25: ...e It is possible to restore these default settings at any time by doing the following 1 Turn the instrument on and make sure it is in the Technician Mode See Section 2 2 4 2 Press the Alarm button on...

Страница 26: ...tivity of the Ex Chek combustible sensor to 50 LEL Do not reset the calibration gas concentration unless you are using a calibration gas concentration that differs from the one that is normally suppli...

Страница 27: ...hanged Chapter 6 Trouble Shooting and Repair 6 1 Replacing Ex Chek sensor The Ex Chek must be calibrated immediately following any sensor change Failure to calibrate the Ex Chek after a sensor change...

Страница 28: ...w instrument to warm back up to room temperature Recharge the batteries and attempt to turn the instrument back on If the instrument still fails to turn on return to factory for repair 6 3 5 No audibl...

Страница 29: ...g materials which similarly protect the instrument and accessories Note The return authorization number must be clearly marked on the outside of the box Prominently showing the return authorization nu...

Страница 30: ...sted Calibration Gases Use of non standard calibration gas and or calibration kit components when calibrating the Ex Chek can lead to dangerously inaccurate readings and may void the standard Biosyste...

Страница 31: ...Must Be Activated By date whichever comes first Mighty Tox 2 Mighty Tox 2 prorated credit is given towards repair or purchase of a new unit 0 6 months of use 100 credit 6 12 months of use 75 credit 12...

Отзывы: