biosystems EX CHEK Скачать руководство пользователя страница 17

15

(3) Insert the end of the sample probe into

the location to be sampled.

(4) Squeeze the aspirator bulb several times

to draw air from the remote location to
the sensor compartment.  Allow one
squeeze of the bulb for every one foot of
sampling hose for the sample to begin
reaching the sensor compartment.  Once
the sample has reached the sensor,
continue to squeeze the bulb at a rate of
once per second for another full minute
until readings stabilize.

Caution:  Hand aspirated remote
sampling only provides continuous gas
readings as long as the bulb is being
squeezed.

2.5  Combustible gas sensor

The Ex Chek contains a single combustible
gas sensor.
It is important to note that catalytic "hot
bead" type combustible sensors require the
presence of oxygen (at least 10 percent by
volume) in order to detect accurately.  A
combustible sensor located in a 100 percent
pure combustible gas or vapor environment
containing no oxygen will be unable to burn
the combustible gas and will produce a
reading of zero percent LEL.

For a detailed discussion of the theory
behind the Ex Chek combustible sensor,
see section 1.3 above.

2.5.1 Detection of missing or

damaged Ex Chek sensor

If the Ex Chek sensor has been
inadvertently removed or has sustained
internal damage the Ex Chek will sound an
audible alarm and flash the visible alarm
LED while displaying “F L T” on the LCD
display. The Ex Chek will stay in alarm
mode until the unit is shut off and an
operational sensor is installed.

Chapter 3. Calibration

The Ex Chek detector has been designed
for easy calibration.  The on / off “MODE”
button is used to enter “Auto-Calibration”
mode, and to automatically make all
calibration adjustments.
It is also possible to manually calibrate the
instrument by using simple push-button
controls located on the instrument keypad.

“One-Button Auto-Cal” procedures are
discussed in Section 3.4.  Manual
calibration procedures are discussed in
Section 3.5.

3.1

Verification of accuracy

  

The accuracy of Ex

Chek instruments should be checked
periodically with known concentration
calibration gas.  Failure to check
accuracy can lead to inaccurate and
potentially dangerous readings.

Verification of accuracy is a two step
procedure.  In the first step the Ex Chek is
taken to an area where the atmosphere is
fresh and the readings are checked.  If the
readings differ from those expected in fresh
air a "zero" adjustment must be made.  The
second step is to make sure the sensor is
accurate by exposing it to known
concentration test gas and noting the
response.  LEL readings are considered
accurate when they are within ±10% of the
expected concentration as given on the
calibration gas cylinder.  If readings are
accurate, there is no need to adjust your gas
detector.  If the readings are inaccurate, the
instrument must be span calibrated before
further use.

  

Always check the

expiration date on calibration gas
cylinder(s) prior to use.  Expired
calibration gas can lead to inaccurate
and potentially dangerous readings.

Содержание EX CHEK

Страница 1: ...0 Reference Manual Ex Chek NiMH Rechargeable Version Combustible Gas Detector 651 South Main Street Middletown CT 06457 800 711 6776 860 344 1079 Fax 860 344 1068 Part number 13 123 Version 2 2...

Страница 2: ...1...

Страница 3: ...IS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED AVERTISSEMENT LIRE ATTENTIVEMENT LES INSTRUCTIONS AVANT DE METTRE EN MARCHE Ex Chek Reference Manual NiMH Version Biosystems Part Number 13 12...

Страница 4: ...t 9 1 9 3 Weather cover 10 1 10 Ex Chek design components 10 1 11 Ex Chek accessories 10 Chapter 2 Basic operation 11 2 1 Overview of Ex Chek operation 11 2 1 1 Operational warnings and cautions 11 2...

Страница 5: ...Chek Advanced Functions 24 5 1 Advanced features overview 24 5 2 Re setting the calibration gas concentration 24 5 3 Setting the security beep interval 24 Chapter 6 Trouble Shooting and Repair 25 6 1...

Страница 6: ...C T4 for instruments powered using DEMKO approved alkaline batteries The Ex Chek is also CE Marked Conformit Europ enne The UL CSA and DEMKO classifications marks indicate compliance with United State...

Страница 7: ...stored and the instrument is once again fit for use 6 Do not reset the calibration gas concentration setpoints in the Ex Chek unless the concentrations of your calibration gas differ from the concentr...

Страница 8: ...6...

Страница 9: ...h a range of zero to one hundred percent explosive The Ex Chek combustible gas sensor is non specific and responds to all combustible gases and vapors Combustible sensors contain two coils of fine wir...

Страница 10: ...n off MODE button once at any time during normal operation causes the backlight to be activated for 10 seconds In the event of a gas alarm the backlight turns on automatically and remains on until the...

Страница 11: ...ght and audible alarm are automatically evaluated and the battery is monitored continuously for proper voltage The Ex Chek offers the following additional options that allow the user to customize the...

Страница 12: ...as control most other functions 6 Protective cap An o ring sealed cap and moisture barrier filter protect the sensor from damage 7 Sensor The top mounted sensor may be easily accessed for replacement...

Страница 13: ...y can lead to inaccurate and potentially dangerous readings 5 A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument that fails calibration ma...

Страница 14: ...appear indicating the current peak alarm set point If the Ex Chek is operating in the Technician mode it will display the following factory installed default alarm set point P E A K A L R M 1 0 This...

Страница 15: ...ve Pressing the MODE button once displays the peak reading of the gas being measured since the instrument was switched on E X 2 0 Pressing the MODE button again returns the display to the first screen...

Страница 16: ...concentration test gas before the instrument is put back into service CAUTION To achieve optimal charge and ensure long battery life of the NIMH battery make sure that charging takes place in an area...

Страница 17: ...l stay in alarm mode until the unit is shut off and an operational sensor is installed Chapter 3 Calibration The Ex Chek detector has been designed for easy calibration The on off MODE button is used...

Страница 18: ...nsitivity is maintained over the life of the sensor 3 1 1 1 Effects of high concentrations of combustible gas on the combustible sensor The accuracy of combustible sensors may also be affected by expo...

Страница 19: ...rument using the procedures discussed in Section 3 4 1 Slip on the calibration adapter as discussed in Section 3 4 2 and flow calibration gas to the sensor Wait for the readings to stabilize Forty fiv...

Страница 20: ...l show the message WAIT ZERO while the adjustment is being completed W A I T Z E R O 6 After successful completion of the zero calibration the display will prompt you with the message SPAN MODE to con...

Страница 21: ...s the concentration shown on the Span Adjust screen as the sensor is adjusted If the concentration does not match it will be necessary to assign a new calibration gas concentration as discussed in Cha...

Страница 22: ...hat the operating mode is changed the LCD screen will briefly indicate the current operating mode 5 Turn the instrument off by pressing the MODE button for three full seconds 6 With the unit turned of...

Страница 23: ...the LCD screen indicates that calibration is complete by showing the message DONE The Ex Chek will then turn itself off D O N E The Cal button must be held down until the screen indicates that span c...

Страница 24: ...e Ex Chek are very conservative in order to maximize worker safety To enter the alarm adjust mode it is necessary to do the following 1 Turn the instrument on and make sure it is in Technician Mode Se...

Страница 25: ...e It is possible to restore these default settings at any time by doing the following 1 Turn the instrument on and make sure it is in the Technician Mode See Section 2 2 4 2 Press the Alarm button on...

Страница 26: ...tivity of the Ex Chek combustible sensor to 50 LEL Do not reset the calibration gas concentration unless you are using a calibration gas concentration that differs from the one that is normally suppli...

Страница 27: ...hanged Chapter 6 Trouble Shooting and Repair 6 1 Replacing Ex Chek sensor The Ex Chek must be calibrated immediately following any sensor change Failure to calibrate the Ex Chek after a sensor change...

Страница 28: ...w instrument to warm back up to room temperature Recharge the batteries and attempt to turn the instrument back on If the instrument still fails to turn on return to factory for repair 6 3 5 No audibl...

Страница 29: ...g materials which similarly protect the instrument and accessories Note The return authorization number must be clearly marked on the outside of the box Prominently showing the return authorization nu...

Страница 30: ...sted Calibration Gases Use of non standard calibration gas and or calibration kit components when calibrating the Ex Chek can lead to dangerously inaccurate readings and may void the standard Biosyste...

Страница 31: ...Must Be Activated By date whichever comes first Mighty Tox 2 Mighty Tox 2 prorated credit is given towards repair or purchase of a new unit 0 6 months of use 100 credit 6 12 months of use 75 credit 12...

Отзывы: