8.2 EMC - Guidance and Manufacturer’s Declaration
The BioSonic Suvi uses RF energy only for its internal function.
Therefore, it's RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Group 1
RF emissions
CISPR 11
RF emissions
CISPR 11
Class B
The BioSonic Suvi is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer's declaration - electromagnetic emissions
Emission test
Compliance
Electromagnetic environment - guidance
The BioSonic Suvi is intended for use in the electromagnetic environment specified below.
The customer or the user of the BioSonic Suvi should assure that it is used in such an environment.
The BioSonic Suvi is intended for use in the electromagnetic environment specified below.
The customer or the user of the BioSonic Suvi should assure that it is used in such an environment.
3 A/m
3 A/m
IEC 61000-4-8
<5 %
U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in U
T
)
for 5 cycles
70 %
U
T
(30 % dip in U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in U
T
)
for 5 sec
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Power frequency
(50/60 Hz)
magnetic field
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
Electrical fast
transient/burst
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
IEC 61000-4-4
Surge
IEC 61000-4-5
<5 %
U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in U
T
)
for 5 cycles
70 %
U
T
(30 % dip in U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in U
T
)
for 5 sec
Guidance and manufacturer's declaration - electromagnetic immunity
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
BioSonic Suvi requires continued
operation during power mains
interruption, it is recommended that the
BioSonic Suvi be powered from an
uninterruptible power supply or battery.
Immunity test
IEC 60601
test level
Compliance level
Voltage dips, short
interruptions and
voltage varaiations
on power supply
input lines
IEC 61000-4-11
Warranty and Declaration of Conformity