Recirculator II User and Service Manual 

 

 

 
 
 

 

 
 
 

Introduction to the Recirculator II 

The Recirculator II is an air purifying device designed to trap and retain airborne particulates. 
Using the latest HEPA (High Efficiency Pariculate Air) filter technology, this unit is able to 
remove 99.97% of all particulate nuclei which exceed 0.3 microns in size. 

Unpacking the Unit 

The Recirculator II is packed fully assembled and tested. When unpacking the unit: 

1.

 

Carefully check the exterior of the packing case for any signs of shipping damage. If 
any damage is discovered, either refuse delivery or contact the carrier.  

2.

 

Carefully remove the unit from the shipping case and stand it upright.  

3.

 

Again check the unit for any signs of shipping damage.  

4.

 

Remove packing reinforcement, unlatch the lid, and remove. 

CAUTION: Care should be taken when working inside the unit. The HEPA 
filters should not be touched, as the filter fabric can easily be nicked or torn, 
which will impair their efficiency. 

5.

 

Check the interior for loose parts or any signs of damage.  

Testing the Unit 

To test the unit: 

1.

 

Re-install the lid and securely fasten both latches, locking with padlocks (not 
supplied) if desired. 

2.

 

Connect the hospital grade plug end of the power cord to a standard wall receptacle 
(110-120 volt, 60 cycle AC electrical outlet). 

3.

 

Turn the unit on by depressing the top (high speed position) of the rocker switch 
located at the bottom rear corner of the unit. 
The blower will start and filtered air will immediately begin to flow from the grilles 
on each side of the unit. 

4.

 

Observe the magnehelic gauge on the front of the unit. 
With the power switch in the upper position as above described (high speed), the 
guage should indicate approximately 0.60-0.65 inches w.c. and be clearly in or below 
the green area on the gauge. This gauge displays the differential in pressure between 
the plenum (inside the unit) and the atmospheric (room) air pressure 

Содержание 9431

Страница 1: ...Biosense Inc 2616 Peartree Lane San Jose CA 95121 800 941 2844 RECIRCULATOR II Part No 9431 User and Service Manual 4th Edition Revised 3 20...

Страница 2: ...Recirculator II User and Service Manual 1...

Страница 3: ...he Recirculator II 4 Servicing the Motor 5 Changing the Filters 6 HEPA Filter Replacement 7 Pre Filter Replacement 8 Cleaning the Recirculator II 8 Appendix A Negative Pressure Adapter Part No 9405 9...

Страница 4: ...A filters should not be touched as the filter fabric can easily be nicked or torn which will impair their efficiency 5 Check the interior for loose parts or any signs of damage Testing the Unit To tes...

Страница 5: ...lator II is equipped with a two speed blower controlled by a dual position switch High Off Low At the High speed upper position the unit will purify about 575 cfm with new clean filters installed At t...

Страница 6: ...tions the motor bearings should be re lubricated every six months with 10 or 20 drops of SAE 10W or 20W non detergent oil ML type or with electric motor oil Refer to diagram at the end of this manual...

Страница 7: ...culator II are an extruded aluminum frame enhanced performance HEPA filter designed to provide approximately 575 cfm when new at a pressure differential of about 0 60 0 65 inches w c When the pressure...

Страница 8: ...old filter a Turn a waste bag inside out and lay it on top of the filter to be replaced b Grasp the top of the filter with the bottom of the bag c Slowly withdraw the filter while rolling the inverted...

Страница 9: ...er the filter 4 Seal the bag and dispose of it properly 5 Insert slide in a new filter noting the direction of airflow arrow should point up and secure the catch NOTE The Recirculator II s performance...

Страница 10: ...ng the Negative Pressure Adapter Unpack the NPA 270 and inspect for damage The two insertion pins contain manufactured bends but should extend parallel with the face of the unit The following are gene...

Страница 11: ...taff HVAC personnel or independent HVAC contractors may now attach appropriate ductwork to the NPA 270 outlet in order to exhaust airflow from the area It should be noted that once fitted with the NPA...

Страница 12: ...ility in all events is limited to and shall not exceed the purchase price paid Prompt Disposition Biosense will make a good faith effort for prompt correction or other adjustment with respect to any p...

Отзывы: