10 

5.  Unit verification 

5.1. 

Checking the calibration.

 

 

If the user have any doubts in measuring precision, he may perform a basic control 
check as follows: 

5.1.1.  No microplate method. 

 

Remove microplate and close the lid. Perform measurement according to p. 4.4. Re-
sults should be between -0.003 and 0.005. 

5.1.2.  Clean microplate method. 

 

Place a clean and empty microplate in the unit, and close the lid. Perform measure-
ment according to p. 4.4. Results should be between 0.030 and 0.060. 

 

Note.

 

For  further  information,  please  consult  with  the  software  operation 
manual. 

5.2. 

Using the verification microplate

 

Precision of the unit is verified by the certified verification plate. Verification certificate 
is enclosed in the standard set and is valid one year from verification date.  

 

All verification procedures must be performed only by qualified and specially trained 
personnel. 

 

Note.

 

Neutral filter absorbance of MULTISKAN

™ verification plates varies 

from plate to plate, so we include an option to set the details of your 
own verification plate. To edit the details, please consult with the soft-
ware operation manual, paragraph 

Editing a verification plate

5.2.1.  Lift the lid and place the verification plate on the sliding platform, then close the lid. 

5.2.2.  To run the certification test, you need to enter the developer mode. To do so, in soft-

ware, open a new experiment, focus on 

the table “Cell, Name, Sample Name, Type” 

(fig. 4) by left clicking on the table area, and press 

Ctrl + N

 

Fig. 4. 

Содержание HiPo MPP-96

Страница 1: ......

Страница 2: ...2...

Страница 3: ...tents 1 Safety Precautions 4 2 General Information 5 3 Getting started 6 4 Operation 7 5 Unit verification 10 6 Specifications 12 7 Maintenance 14 8 Warranty and Claims 14 9 EU Declaration of Conformi...

Страница 4: ...power switch and external power supply are easily accessible during use Disconnect the unit from the external power supply before moving This equipment is connected and controlled by PC Please ensure...

Страница 5: ...BestFit function for the selection of the best calibration curve Multiplex analysis up to 7 different tests on the same plate Qualitative assays the ability to install up to 8 types of controls weak p...

Страница 6: ...pce Operating manual 1 copy Certificate of verification 1 copy Optional accessories Light filters custom 400 700 nm on request Verification microplate set on request Verification microplate set 3 3 Se...

Страница 7: ...A1 is placed next to the A1 label tag fig 3 1 Close the lid Note Measurements can be affected by uneven or heterogeneous well contents Visually inspect the plate for foaming opaqueness bubbles or part...

Страница 8: ...seconds 4 4 4 Press Start button The Measurement LED light fig 1 2 turns on Attention Do not open the lid during measurements 4 5 Measurement processing After an 8 second measurement the software aut...

Страница 9: ...mat press XLS 96 well button 4 9 After completing the measurements open the lid and remove the microplate then close the lid Attention Do not leave the microplate in the unit 4 10 After finishing the...

Страница 10: ...the certified verification plate Verification certificate is enclosed in the standard set and is valid one year from verification date All verification procedures must be performed only by qualified a...

Страница 11: ...channels to test fig 5 4 and the number of measurements to perform fig 5 2 We recommend a minimum of 20 measure ments Press the Start button fig 5 3 5 2 4 When measurements are finished press the Sho...

Страница 12: ...channel 1 6 9 Optical density measurement range 0 4 3 OD 6 10 Resolution 0 001 OD 6 11 Accuracy at 405 450 492 620 nm 0 000 2 000 OD 0 5 0 010 OD 2 000 3 000 OD 1 0 010 OD 6 12 Precision reproducibili...

Страница 13: ...imum spectral power are not equal Table 1 96 well plate dimensions 96 well plate Dimension mm Plate height 14 35 Plate length 85 48 Plate width 127 76 First well position X 14 38 First well location Y...

Страница 14: ...ments of Specifications provided the Customer follows the operation storage and transporta tion instructions 8 2 The warranted service life of the unit from the date of its delivery to the Customer is...

Страница 15: ...ctive 2014 30 EU LVD Directive 2014 35 EU RoHS2 2011 65 EU WEEE 2012 19 EU Applicable Standards LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Gen...

Страница 16: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 1 07 January 2017...

Отзывы: