6.  Maintenance 

6.1. 

If the unit requires maintenance, disconnect the unit from the electric circuit and con-
tact Biosan or your local Biosan representative. 

6.2. 

All maintenance and repair operations must be performed only by qualified and spe-
cially trained personnel. 

6.3. 

Care and Cleaning  

6.3.1.  Standard ethanol (75%) or other cleaning agents recommended for cleaning and de-

contamination of the stainless steel surfaces. 

6.3.2.  For decontamination, it is recommended to use a special DNA/RNA removing solu-

tion (e.g. DNA-Exitus Plus

TM

, RNase-Exitus Plus

TM

). 

6.4. 

Fuse replacement.  

 

Disconnect the power cord from electric circuit. Disconnect the power cable from the 
socket on the rear of the shaker-incubator. Open the fuse holder. Check and replace 
with the correct fuse if necessary: for 230 V, M 3.15 A or for 120 V, M 5.0 A (type 

M

 

- time lag: 

M

edium).  

 

Fig. 2. Fuse replacement 

 

 

Содержание ES-20/60

Страница 1: ......

Страница 2: ...2 Contents 1 Safety precautions 3 2 General Information 4 3 Getting started 4 4 Operation 6 5 Specifications 8 6 Maintenance 9 7 Warranty and Claims 10 8 EU Declaration of Conformity 11...

Страница 3: ...and plug are easily accessible during use Disconnect the unit from the electric circuit before moving Disconnect the plug from power socket to turn off the unit If liquid penetrates into the unit dis...

Страница 4: ...on and temperature regulation PID control decrease the energy consumption and make the shaker incubator highly energy efficient despite its relatively large size 3 Getting started 3 1 Unpacking Remove...

Страница 5: ...ort its weight remove protective film from the display plug the power cord into the socket on the rear and position the unit so that there is easy access to the power switch and plug 3 4 Platform inst...

Страница 6: ...s required The set parameters can also be changed during oper ation 4 3 1 Setting time Time Using the and Time keys Fig 1 1 set the required work ing time interval in hours and minutes increment 1 min...

Страница 7: ...g the RPM Run Stop key will start continuous operationof the unit with countdown timer in the lower line and indication OFF in the upper line until the RPM Run Stop key is pressed again Caution At the...

Страница 8: ...ng voltage power consumption 230 V 50 60 Hz 450 W 2 A or 120 V 50 60 Hz 450 W 4 5 A Weight 41 1 kg Optional ac cessories Description Catalogue number P 30 100 Platform with clamps for 30 flasks of 100...

Страница 9: ...ning agents recommended for cleaning and de contamination of the stainless steel surfaces 6 3 2 For decontamination it is recommended to use a special DNA RNA removing solu tion e g DNA Exitus PlusTM...

Страница 10: ...ess To obtain the claim form visit section Technical support on our website at link be low 7 5 Extended warranty For ES 20 60 a Premium class model one year of extended warranty is available free of c...

Страница 11: ...laboratory use EMC requirements General requirements LVS EN 61010 1 2011 Safety requirements for electrical equipment for measurement control and laboratory use General requirements LVS EN 61010 2 010...

Страница 12: ......

Отзывы: