Page 9
V1.6.3
Warning!
Do not use the BioMagPure 12 Plus system to a location where it is wet or can be
splashed with water. It may cause a device failure, fire, or electric shock.
When relocating the BioMagPure 12 Plus system, disconnect the plug from the
outlet first. If the power cable is damaged, this may cause a device failure, fire,
injury, or electric shock.
Caution!
Do not use the BioMagPure 12 Plus system in an unstable place such as a slanted
surface or a place subject to vibrations. It may cause injury or device failure.
Do not use the BioMagPure 12 Plus system in direct sunlight or close to a heating
device. It may shorten the life of the BioMagPure 12 Plus system, or cause a trou-
ble.
Do not open the maintenance door while performing experiment.
Do not open the drawer while performing experiment.
3.5.
Operating Conditions
Items
Conditions
Temperature (°C)
During operation
15
– 40
During down time
0
– 55
Humidity (RH)
During operation
30
– 80
During down time
10
– 80
Max. wet bulb temperature (°C)
During operation
29 (non condensing)
During down time
29 (non condensing)
Temperature gradient (°C/hr)
12 or less (non condensing)
Humidity gradient (RH/day)
30 or less (non condensing)
Altitude, operating (m)
1600 or less
3.6.
Unpacking of the BioMagPure 12 Plus system
3.6.1. Open the packing box and take out the instrument and related
accessories.
Note.
The BioMagPure 12 Plus system has a weight of more than 60 kg. It should be lift
and moved by two persons. Hold the moving handler of the instrument from two
sides to move it out from the box. Do not lift by the outer covering, front panel or
the door.
Caution!
Improper handling of the movement of the BioMagPure 12 Plus system will lead to
instrument damages.
3.6.2. For correct and safe use of the BioMagPure 12 Plus, install it in a location that is close to
electrical outlet and has
enough space for installation and operation of the mains switch.
Note.
Keep the shipping box and stuffing materials. They will be needed again
when
transporting the BioMagPure 12 Plus system.
3.6.3. Connect Barcode Reader with the Instrument, see
3.7
.