Page 16
V1.6.3
5.
Cleaning and maintainance
5.1.
Two types of maintenance are performed on the BioMagPure 12 Plus system Instrument as
listed in the table. For details on each type of maintenance, see below:
Maintenance Type
Performed by
Schedule
Cleaning of Sample Tray
User
After each use
Cleaning of instrument body (outside)
Bi-weekly
Cleaning of instrument body (inside)
Preventive
Service Engineer
Annually
5.2.
Clean the Sample Tray with mild detergent and rinse with deionized water. Allow the parts to
dry before use. Clean and disinfect the platform surface by wiping with deionized water followed
by 75% ethanol.
5.3.
Clean the instrument body by removing dust gently with a dry, soft cloth. If the outside of the
BioMagPure 12 Plus system is heavily soiled, or if any samples that may cause infection are
adhered to the outside of the BioMagPure 12 Plus system, wipe with a soft paper tissue,
soaked with 0.5% sodium hypochlorite solution or ethanol.