Due to unstable conditions during
measurements, it is not possible to calculate
an average result.
Avoid unnecessary movement and
talking.
Low battery
Replace batteries.
Cuff pressure is over 300 mmHg or pulse is
over 200 beats per minute.
Ensure that the cuff is worn correctly
and measure again. Avoid movement
or talking when the cuff is being
inflated.
Pulse below 40 is detected.
Ensure that the cuff is worn correctly.
If problems occur when using the device the following points should be checked, and if necessary, the
corresponding measures should be taken.
Malfunction
Remedy
The display remains empty when the device is
switched on. The batteries are inserted.
1. Check batteries for correct polarity.
2. If the display is unusual, re-insert the batteries
or exchange them for new ones.
The pressure does not rise even though the pump
is running.
Check the connection of the cuff tube and
connect properly if necessary.
The device frequently fails to measure the blood
pressure values, or the values measured are too
low or too high.
1. Check the positioning of the cuff.
2. Measure the blood pressure again, ensuring
that you have remained motionless for
a sufficient amount of time to ensure an
accurate reading.
Every measurement produces varying results
although the instrument functions normally and
the values displayed are normal.
Note that blood pressure fluctuates continuously;
therefore measurements will show some
variability.
Blood pressure values measured differ from those
measured by the doctor.
Record daily measurements for consultation with
your doctor.
Note: Individuals visiting their doctor
frequently experience anxiety which can
result in a higher blood pressure reading
than at home.
For assistance call the BIOS Medical Hotline: 1-866-536-2289.