3.9 Viewing Previously Recorded Values
The blood pressure monitor automatically stores your measurements with time and
date. It can store up to 99 measurements for each of the 2 users. When more than 99
measurements are made, the oldest readings are deleted for that particular user to
make space for the new ones.
To view the previously stored values, press the button. The last measurement is
displayed. The date and time of the measurement are also displayed after the reading.
Press the button repeatedly to view all the measurements that are
recorded on the device.
3.10 Clearing All Values
If you are sure that you want to permanently remove all stored values, hold down the
button (the instrument must have been switched off before hand) until the “CL”
appears and then release the button. If you do not want to clear the values, press
the button. To permanently clear the memory, press the button while “CL” is
flashing. Individual values cannot be cleared.
3.11 Discontinue a Measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g. the
patient feels unwell), the button can be pressed at any time. The device then
immediately lowers the cuff pressure automatically.
4. Error Messages / Malfunctions
If an error occurs during a measurement, a long beep followed by two short beeps is generated and the LCD
display the corresponding error code.
Error
Possible Cause
Remedy
Err 1
No pulse has been detected.
Ensure that the cuff is being worn correctly,
and that you have your arm at the heart level.
Err 2
Unnatural pressure impulses influence the
measurement result. Reason: The arm was moved
during the measurement.
Avoid unnecessary movement or talking.
Err 3
The inflation of the cuff takes too long. The cuff is
not correctly seated.
Ensure that the cuff is being worn correctly.
Err 5
The difference between systolic pressure and
diastolic pressure is too far away from acceptable
and reasonable range.
Ensure that the cuff is being worn correctly
and that you have been inactive for a sufficient
time before making the measurement.
Err 6
Due to unstable conditions during measurements,
it is not possible to calculate an average result.
Avoid unnecessary movement and talking.
Low battery
Replace batteries.