www.biopac.com
BIOPAC Hardware |
OUT SERIES
| Page 4 - 4
Updated: 1.6.2014
OUT103 LED CABLE
Use this LED cable to synchronize a light flash. The 3 meter cable
makes it easy to use the LED for a variety of protocols. Terminates
for connection to Analog OUT 0/1 and includes adapter for
connection to Digital I/O.
Media synchronization
- Windows only -
Acq
Knowledge
4.1 and above
MP150 AND UIM100C SETUP
a.
Connect the OUT103 2 mm pin adapter to the 3.5 mm plug on the OUT103 cable.
b.
Connect the red OUT103 2 mm pin to a Digital I/O channel on the rear of the UIM100C and the
black pin to GND D on the rear of the UIM100C.
c.
Use MP150 > Set Up Channels to acquire and plot the Digital I/O channel the OUT103 is connected
to.
d.
Set MP150 > Show Manual Control
Set for ‘Output’.
Enable the 'Set immediately' option.
Click the Digital I/O channel the OUT103 was connected to to toggle between 0 and 1.
If necessary, click the 'Set' button to update the manual control and output a digital pulse.
MP36R SETUP - ADDITIONAL ITEMS REQUIRED
a.
Connect an OUT3 (BNC adapter) to the 'Analog Out' port on the rear of the MP36R.
b.
Connect a BSLCBL6 (interface cable: BNC to 3.5 mm) to the OUT3.
c.
Connect the OUT103 3.5 mm plug to the BSLCBL6 3.5 mm socket.
d.
Set MP36 > Output Control 'Low Voltage Stim' option
Set Pulse width to 100 msec.
Set Pulse level to 5 volts – set Reference Channel to any digital channel.
Click the D’ON’ button to output a digital pulse.
CALIBRATION
The OUT series does not require calibration.
40HP MONAURAL HEADPHONES
These monaural headphones are used with Biopac Science Lab MP40 and
Biopac Student Lab MP45 for stimulus response experiments and to listen
to EMG signals. The 40HP is a wide-response, high-efficiency headphone.
40HP SPECIFICATIONS
Cable Length:
5 meters
Connector Type:
3.5 mm phone plug
HARDWARE GUIDE