background image

6

 - 

BIOPAC Systems, Inc. 

Visit 

www.biopac.com

 for the latest product updates 

SPECIFICATIONS 

Specification 

MP35 Unit 

MP30 Unit 

Front Panel

 

 

 

ELECTRODE CHECKER 

Resistance Range (Vin+ and 
Vin- to GND) 

0-100 K

 

 

0-100 K

 

ANALOG INPUTS  
Number of channels 

4 isolated (front panel CH 1–CH 4), 
2 unisolated (auxillary) 

4 (front panel CH 1–CH 4) 
 

SAMPLE RATE 

Maximum 
Minimum 
Trigger Input 
Threshold 

 
100K samples/second 
1 samples/second 
Analog or digital channel 
Adjustable threshold; Positive or 
Negative Trigger 

 
2K s/s (8K aggregate on four ch.) 
1 samples/second 
CH 4 input only 
Adjustable threshold; Positive or 
Negative Threshold 

A/D resolution (before digital 
filtering) 

24-bit  

10-bit 

Signal to noise ratio 

> 90 dB (nominal) 

> 90 dB 

Voltage resolution Gain 
dependent 

1.192 microvolts /bit (Gain 10) to 
0.024 nanovolts /bit (Gain 50,000) 

0.400 microvolts/bit (Gain 100) to 
0.200 millivolts/bit (Gain 25,000) 

Input voltage range (Gain 
dependent) 

400 microvolts to 2.0 Volts p-p 

4.0 millivolts to 0.2 Volts p-p 

Input accuracy 

±0.01% of Full Scale Range (FSR) 

±0.05% FSR 

Input protection; current limited 

± 1 mA/V 

± 1 mA/V 

Maximum Input Voltage 
(between Vin+ and Vin-) 

2V p-p  

130mV p-p 

Differential Input Impedance 
(between Vin+ and Vin-) 

2 M

  

2 M

 

Filters (automatic or user 
adjustable) 

3 two-pole IIR digital filters per 
channel  

3 two-pole IIR digital filters per 
channel 

Common Mode Input 
Impedance (between Vin+/Vin- 
and GND) 
  DC 
  AC (50/60 Hz) 

 
 
 
11 M

 

1,000 M

  

 
 
 
11 M

 

1,000 M

 

Gain ranges (automatic preset 
or user adjustable) 

10 – 50,000  

100 – 50,000 

Baseline adjustment 
(automatic or user adjustable) 

Gains 10, 20, and 50: ±100mV 
Gains 100 to 50,000: ±10mV 

±10mV all Gains 

Electrode offset potential 
tolerance 

Gains 10, 20, and 50: ±2V 
Gains 100, 200, 500: ±200mV 
Gains 1,000 to 50,000: ±80mV 

±70 mV all Gains 

Back Panel

 

 

 

ANALOG OUTPUT 
  Number of channels 
  D/A resolution 
  Accuracy 
  Output impedance 
  Output voltage 
  Output drive current

 

 

12 bits 
±0.0125% of FSR 
50

 

0 - 4.096 V 
±10 mA maximum 

 

8 bits 
±0.2% of FSR 
50

 

0 - 5.000 V 
±100 mA maximum 
 

Содержание MP30

Страница 1: ...ted data acquisition unit designed for biophysical measurements Exercise extreme caution when applying electrodes and taking bioelectric measurements while using the Biopac Student Lab with other external equipment that also uses electrodes or transducers that may make electrical contact with the Subject Always assume that currents can flow between any electrodes or electrical contact points In ca...

Страница 2: ...uments On partial Turns MP35 on assuming AC300A power adapter is powered by the mains Off partial Turns MP35 off if but AC300A power adapter remains powered by the mains Direct current Direct current output USB USB port COMPLIANCE SAFETY The MP35 satisfies the Medical Safety Test Standards affiliated with IEC60601 1 The MP35 is designated as Class I Type BF medical equipment EMC The MP35 satisfies...

Страница 3: ... of the MP35 30 has an electrode check port four analog input ports and two status indicators Electrode Check The Electrode Check port is a diagnostic tool used with the BSL PRO software to determine if the electrodes are properly attached to the subject See page Error Bookmark not defined Input ports CH 1 CH 2 CH 3 and CH 4 The inputs on the MP35 30 acquisition unit are referred to as Channels Th...

Страница 4: ...SB connector Should only be used to connect the MP35 to a PC or Macintosh Serial port MP30 only The MP30 connects to the computer via a serial port located just below the word Serial Uses a standard MINI DIN 8 connector Should only be used to connect the MP30 to a PC with ISA or PCMCIA card or Macintosh Headphone Output MP35 only Accepts a standard 1 4 or 6 3mm stereo headphone jack I O Port MP35 ...

Страница 5: ... from shorts on its power analog and digital I O lines The MP35 uses a 1 0 amp fast blow fuse and the MP30 uses a 2 0 amp fast blow fuse To remove the fuse use a screwdriver to remove the fuse cover located below the word Fuse Power switch ON position powers up the MP35 30 OFF position cuts the flow of power to the MP35 30 CLEANING PROCEDURES Be sure to unplug the power supply from the MP35 30 bef...

Страница 6: ...put voltage range Gain dependent 400 microvolts to 2 0 Volts p p 4 0 millivolts to 0 2 Volts p p Input accuracy 0 01 of Full Scale Range FSR 0 05 FSR Input protection current limited 1 mA V 1 mA V Maximum Input Voltage between Vin and Vin 2V p p 130mV p p Differential Input Impedance between Vin and Vin 2 MΩ 2 MΩ Filters automatic or user adjustable 3 two pole IIR digital filters per channel 3 two...

Страница 7: ...ble outputs N A MP35 only TRIGGER MP35 only TTL compatible input and synchronization port N A MP35 only DC INPUT Power input requires 12 VDC 1 Amp Use the AC300A 12 VDC 1 25 Amp power supply adapter to connect to any mains rated as 100 250 VAC 50 60Hz 40VA Power input requires 12 VDC 1 Amp Use the AC300A 12 VDC 1 25 Amp power supply adapter to connect to any mains rated as 100 250 VAC 50 60Hz 40VA...

Страница 8: ...UB Pin MP35 MP30 1 Buffered AC output Buffered AC output Z out 2 200 µF Cap Z out 2 200 µF Cap V out range MP35 2 0 V MP30 2 5 V 2 Buffered DC output Buffered DC output Z out 50 Ω Z out 50 Ω V out range MP35 0 to 4 096 V MP30 0 to 5 V 3 GND GND 4 5 0 V 100 mA max 7 5 V 100 mA max 5 Buffered digital output Unbuffered DC output Z out 1 kΩ Z out 1 kΩ V out range 0 to 5 V V out range 0 to 5 V 6 12 V 1...

Страница 9: ...utput 1 2 Digital Output 2 3 Digital Output 3 4 Digital Output 4 5 GND Unisolated 6 GND Unisolated 7 RS 232 RX 8 5 V Unisolated 9 I2 C SDA 10 Digital Input 1 11 Digital Input 2 12 Digital Input 3 13 Digital Input 4 14 Digital Output 5 15 Digital Output 6 16 Digital Output 7 17 Digital Output 8 18 Analog Input Right 19 Analog Input Left 20 RS 232 TX 21 I2 C SCL 22 Digital Input 5 23 Digital Input 6...

Отзывы: