Alternative Site Testing

Alternative Site Testing

CAUTION

- Consult your healthcare professional before sampling from your palm or forearm. 
- Alternate site testing should only be done during steady-state time (when glucose is not 

changing  rapidly), fingertip can show the rapid change in glucose faster than palm or 
forearm samples. 

- Only the fingertip should be used when glucose levels changing rapidly conditions are as 

following: after drinking, after a meal, after insulin injection or exercise, during illness, 
during times of stress, you think your blood sugar is low, or the results do not agree with 
the way you feel.

- DO NOT test on the palm or forearm if you are testing for insulin dose calculations or 

hypoglycemia (Low blood glucose). 

- AST result should not be used to calibrate continuous glucose monitors (CGMS).
- Use the clear cap provided with RIGHTEST Lancing Device when testing sites other than 

fingertip. The normal lancing device cap for your palm or forearm, the blood sample may 
not be sufficient for meter operation.

Alternative site testing: palm or forearm blood sampling 

- To perform alternative site testing, install the clear cap for 

your RIGHTEST Lancing Device. (For detailed information, 
refer to the instruction manual for RIGHTEST  Lancing 
Device).

- To increase the blood flow, massage the intended puncture 

area of your

 

palm or forearm for a few seconds. 

-

 

Immediately after massaging the intended puncture area, 

press and hold the lancing device with the clear cap 
against palm or forearm.   

- Press the release button. 

Palm

Forearm

- Keep pressing the lancing device against your palm or forearm and gradually increase pressure 

for a few seconds until the blood sample size is sufficient. DO NOT test on the palm or forearm if 
you are testing for hypoglycemia (Low blood glucose).

Releasing 

button 

Pressing for

a few seconds

Содержание RIGHTEST Max

Страница 1: ...onitoring System BIONIME CORPORATION No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E mail info bionime com Made in Taiwan BI...

Страница 2: ......

Страница 3: ...r concerns please contact your local Bionime Customer Service or email to rightest bionime com RIGHTEST Max Blood Glucose Monitoring System is intended to be used for the quantitative measurement of g...

Страница 4: ...Blood Glucose Monitoring System is NOT intended for diagnosis of diabetes mellitus If RIGHTEST Meter and Test Strips are exposed to temperature environments out of range for the meter below 6 C 43 F o...

Страница 5: ...betes mellitus These test strips should not be used with meters to test critically ill patients Alternative site testing AST should only be performed during steady state times when glucose is not chan...

Страница 6: ...dy for Testing Performing a Blood Glucose Test Setting Markers Recalling Test Results Recalling Average Test Results Alternative Site Testing View Window Appearance Understanding Test Results and Mess...

Страница 7: ...Documents Getting Started Guide User s Manual Log Book Warranty Card Emergency Card 2 RIGHTEST Max Blood Glucose Meter with battery installed 3 RIGHTEST Max Blood Glucose Test strips 10 or 25 pcs wit...

Страница 8: ...lood Glucose Meter Up Button Down Button Main Button Display Window Test Strip Port RIGHTEST Max Blood Glucose Meter Battery Cover Do not put battery here Meter Battery Pen press to open the Cover Met...

Страница 9: ...is low or must be replaced Indicates a meal marker with test result Indicates that the Bluetooth is enabled meter Bluetooth only Indicates a test result stored in memory Indicates a testing result is...

Страница 10: ...C 39 86 F and with 90 relative humidity Do not expose to direct sunlight or heat If RIGHTEST Meter and Test Strips are exposed to a temperature environments out of range for the meter below 6 C 43 F o...

Страница 11: ...enter Self Testing Mode automatically when replacing the battery all symbols will appear on the screen 5 Press any button to exit the self test and enter Setting Mode 6 You must set the time and date...

Страница 12: ...button to confirm each selection After confirming all of the settings you will return to the Time Mode After pressing the main button for 3 seconds the will turn off its screen Please meter keep press...

Страница 13: ...links press the Up or Down button to select the current hour and press the main button to conform the setup 6 Minute setting When the minute blinks press the Up or Down button to select the current mi...

Страница 14: ...feature reminds the user when to perform a test There are 8 sets of alarm available per day The meter will beep when the alarm is triggered and it will be triggered up to 3 times Insert the test stri...

Страница 15: ...rip as the picture shown 3 Insert RIGHTEST Blood Glucose Test Strip into test strip port until it clicks and firmly stops Turning On Off the Meter 1 How to turn on RIGHTEST Blood Glucose Meter Press t...

Страница 16: ...e Sterile lancet Alcohol swab optional Plunger Hub Lancet carrier Release button Depth adjustable cap Removing RIGHTEST Blood Glucose Test Strip 1 Hold RIGHTEST Blood Glucose Test Strip as shown 2 Rot...

Страница 17: ...able lancet 6 Replace the depth adjustable cap 1 3 4 5 6 2 7 The adjustable cap with 7 depth levels allows you to select the depth of penetration by rotating the cap until the preferable depth display...

Страница 18: ...finishing the test 19 15 16 17 18 Performing a Blood Glucose Test 14c 14b 13 12b 12a 33 32 12 After inserting a test strip the icon will flash on the screen Pressing the Up or Down button to choose b...

Страница 19: ...ning Always keep the metal contacts of the test strip port clean If any dust or dirt is present please clean with a small soft brush All parts of this kit are considered biohazards and can potentially...

Страница 20: ...ease the button to stop at any particular test result The RIGHTEST Max Blood Glucose Meter is able to automatically store a maximum of 500 test results with time and date If your meter has stored 500...

Страница 21: ...hable in the Memory Mode The CS test result will show with a blinking icon in the Memory Mode The Lo Hi results control solution test results and test results under abnormal temperature 6 C 43 F 44 C...

Страница 22: ...se the clear cap provided with RIGHTEST Lancing Device when testing sites other than fingertip The normal lancing device cap for your palm or forearm the blood sample may not be sufficient for meter o...

Страница 23: ...ips and lancets View Window Appearance Make sure your blood sample covers the whole area of the view window to get an accurate test result An insufficient blood sample will result in an error message...

Страница 24: ...r RIGHTEST Blood Glucose Monitoring System is working properly The meter displays results between 10 and 600 mg dL 0 6 and 33 3 mmol L If the test result is below 10 mg dL 0 6 mmol L will appear on th...

Страница 25: ...12 About Quality Control Testing When should a Quality Control Test Be Performed To ensure that your RIGHTEST Blood Glucose Meter and RIGHTEST Blood Glucose Test Strip are working properly To confirm...

Страница 26: ...ht leak into the bottle of control solution and may cause the degeneration of the control solution This could contaminate the meter via the test strip port Keep the test strip port clean and dry Clean...

Страница 27: ...disinfecting wipe has been tested and may be used to clean and disinfect the meter and lancing device Maintain the Products Understanding Control Test Results Your control solution test results should...

Страница 28: ...ll blood and body fluids should be cleaned from surface before performing the disinfecting procedure 3 Allow the surface to remain wet for 2 minute 4 Allow to air dry Maintain the Products Maintain th...

Страница 29: ...2 min Cleaning and Disinfecting frequency at least once a week To clean the lancing device 1 Thoroughly wipe the entire surface of the lancing device with disinfecting wipes listed above to clean any...

Страница 30: ...u are using RIGHTEST Blood Glucose Test Strip Sampling Error Please do not apply the blood to the sample port of the test strip before the meter displays Discard the test strip If the meter shows Temp...

Страница 31: ...p Er2 The meter has malfunctioned Do a Quality Control Test or reinstall the battery to check if the meter works properly Er3 The signal transmission is disrupted Repeat the test using a new test stri...

Страница 32: ...40 2 mm 2 3 10 60 C 14 140 F Refer to RIGHTEST Max test strip insert Specification Dehydrogenase Electrochemical Sensor 10 600 mg dL 0 6 33 3 mmol L 5 seconds 500 blood glucose test results with date...

Страница 33: ...n materials and workmanship for five years from the date of purchase This warranty does not apply to the performance of a RIGHTES Max Blood Glucose Meter that has been altered misused tampered with or...

Страница 34: ...Medical Device Directive 93 42 EEC EC Rep Bionime GmbH Tramstrasse 16 9442 Berneck Switzerland Disposable Sterile Lancets Manufacture information is as package printed Expected Glucose Values Referen...