CardioCare2000 Operation Manual
2) Regular Inspection
In common with all medical equipment, it is recommended that a safety check of EKG2000
should be carried out once a year. Refer to Service manual offered by Bionet Co., Ltd. on the
extent of the check.
3) Troubleshooting
1. In case that buzzer alarms for 1 second during printing and following message appears on
the LCD ;
This indicates that printer run out of printing paper. Operate again after replacing printing
paper.
2. In case that nothing can be printed out or printout fades ; Top cover of the printer is not
closed completely. Operate again after closing firmly
3. In case that buzzer alarms three times in series when system is on battery power, following
message appears on the LCD.:
This indicates that the battery is almost discharged. Turn the system power off and connect
the ac power source. Then turn the system power on again.
.
4. In case that ECGs is interfered with power noise;
First, move the menu to FLT
AC and check if AC is set to 60Hz..
If ECGs is still interfered even if ac filter is set to 60Hz, connect the potential equalization
connector with the common ground conductor. Do not connect the potential equalization
conductor with the ac power system ground. The metallic frame of the patient’s bed or another
metallic object connected with building can be used for grounding
.
* RECORD *
Paper empty!
* Low Battery *
Power off
Содержание CardioCare 2000
Страница 1: ...CardioCare 2000 OPERATION MANUAL Ver 2 41 www ebionet com ...
Страница 50: ...50 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 3ch 1rhy ...
Страница 51: ...51 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 6ch 1rhy ...
Страница 52: ...52 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 12ch rhy ...