47 / 61
Ver. 2.41
CardioCare2000 Operation Manual
4) Record mode output
In record mode output, ECGs is recorded on the memory first and displayed according to the
selected signal level, output display speed, channel form after applying the selected filter and
calculating the measurement parameter including heart rate, PR interval, QRS duration, QT/QTc,
P-R-T axes of the recorded ECGs.
10 seconds ECG recording
Press
key when channel form is set to 3ch+1rhy, 6ch+1rhy, 12ch rhy
Then, the system starts recording data for 10 seconds on the memory with the following
message indicates recording time on the LCD.
After recording data for 10 seconds, the system starts applying the selected filter and calculating
the measurement parameter of the recorded ECGs with the following message on the LCD.
After data processing , the system starts printing with the following message on the LCD.
Press
key to exit during recording or printing data. Then, system stops recording or
printing and displays system setting after following message for 1 second on the LCD.
* RECORD(10s) *
10sec
* RECORD(10s) *
Processing
* RECORD(10s) *
Printing
* RECORD(10s) *
Aborted!