64
BM3Vet Pro
User’s Manual
moving.
Note
⚫
Make sure that the contact area of the disposable electrode is not dry to maintain a
good connection between the electrode and the skin.
⚫
If you suspect that the disposable electrode is in poor contact, replace it immediately
with a new electrode. Otherwise, the contact impedance of the skin and electrode will
increase, and the correct ECG signal will not be obtained.
⚫
If the contact condition gets worse before expiration date on the packaging, replace
with a new one.
⚫
To get a stable ECG waveform, rub the skin with gel or benzoin tincture.
ECG Precaution
Caution
⚫
ECG diagnostic interpretation should be performed by a veterinarian or an
individual trained under the supervision of a veterinarian.
⚫
Use caution with potential source equipment as they may interfere with ECG
monitoring.
⚫
Animals prone to epilepsy should not rely entirely on ECG. Electrical waves of non-
cardiac sources such as seizures can interfere with the detection of certain
arrhythmias.
Warning
CABLES
—
Route all cables away from animal's throat to avoid possible strangulation.
CONDUCTIVE CONNECTIONS
—
Extreme care must be exercised when applying medical
electrical equipment. Many parts of the human/machine circuit are conductive, such as the
animal, connectors, electrodes, transducers. It is very important that these conductive parts
do not come into contact with other grounded, conductive parts when connected to the
isolated animal input of the device. Such contact would bridge the animal's isolation and
Содержание BM3Vet Pro
Страница 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Страница 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Страница 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...