9. DEFAULT SETTING VALUE
1. HORSE Mode
Alarm level
High
Medium
Low
Message
SpO
2
- %
0
SpO
2
- Rate
0
NIBP
0
T(
ْ ْ
ْ
C/
ْ ْ
F)
0
Check probe
0
Probe Off
0
Low Battery
0
Parameter Limits
Low
High
NIBP-S(mmHg)
80
200
NIBP-M(mmHg)
50
170
NIBP-D(mmHg)
30
150
SpO
2
- %
90
100
SpO
2
- Rate
60
150
T(
ْ ْ
ْ
C/
ْ ْ
ْ
F )
36.0
℃
/ 96.8
℉
40.0
℃
/ 104.0
℉
Display Defaults
Animal Type
HORSE
NIBP Auto
Off
NIBP Cuff Size
LARGE
LARGE Cuff Pressure
170mmHg
MEDI. Cuff Pressure
140mmHg
SMALL Cuff Pressure
120mmHg
NIBP Limit Type
Systolic
NIBP Units
mmHg /kPa
SpO
2
Check Probe
Message Alarm
Pulse Volume
Off
Units for Height
m/inch
Units for Weight
Kg/LBS
Temperature Units
ْ ْ
C/
℉