ExiPrep
TM
16 Plus
Fully Automated Nucleic Acid Extraction System
-13-
※
Setup process according to the CHECK LIST
8a. Insert Buffer Cartridge
②
on the Base plate.
¾
Make sure the Buffer Cartridge fits snugly and exactly into its place.
Misplacement of the Buffer Cartridge may result in instrument break
down and malfunction.
8b. Insert Buffer Cartridge
①
on the Base plate.
¾
Make sure the Buffer Cartridge fits snugly and exactly into its place.
Misplacement of the Buffer Cartridge may result in instrument break
down and malfunction.
8c. Place the Elution tube rack on the Base plate.
¾
Make sure the direction and location of the Elution tube rack is correct.
8d. Place the Disposable tip rack on the Base plate.
¾
Make sure the direction and location of the Disposable tip rack is
correct.
8e. Place the Waste tray into the gap between Buffer Cartridges
①
and
②
.
8f. Push the Base Plate in completely and close the door.
¾
Press the ‘
ok
’ button to complete.