Treatment Completion
Therapeutic treatment should not be suspended until fusion occurs or until such time
as a determination is made by the prescribing physician. The device is programmed to
deliver 270 continuous days of therapeutic treatment and automatically discontinues
operation after the 270 days.
Patient Compliance Monitoring
The Biomet
®
SpinalPak
®
Non-invasive Spine Fusion Stimulator System contains embedded software
which allows the display of patient specific history data including usage and therapeutic treatment
times. This data may be downloaded to a personal computer for viewing, storage and/or print out via
the use of Biomet Compliance Data Download Software. Please call your local Biomet representative
to obtain more information.
Ordering Information
To order supplies, contact Biomet. See page 2 “Important Safeguards” for contact information.
The following information is necessary to expedite any inquiry:
• Patient name
• Physician name
• Address to send replacement parts (patient home, MD office, etc.)
Electrode Instructions for Use
Do not open outer packet until ready to use.
1) Tear open packet.
2) Remove electrode from clear plastic backing liner.
3) Wet finger with tap water and moisten entire gel area.
4) Place electrode on skin.
5) Connect electrode to electrode lead.
Renewal
1) With continuous use, electrodes may dry out.
2) To renew, wet finger with tap water and moisten entire gel area.
3) Reapply electrode to skin.
Store in a cool place