TG
RADIENT
Manual __________________________________________________________ 20
6 Further programming options
Temp[°C]
time
ß
#
gradient[°C]
opt
à
1: 95.0
0h 5m 0s
2: 95.0
0h 0m30s
3: 55.0
0h 0m30s
4: 72.0
0h 0m30s
2
A ?
B header
C pgm OK
D enter
Note:
To program special parameters you can scroll the display to the right.
Move the cursor to the right side of the display that is labeled with opt
à
.
The following screen will appear:
Main funct.
dT[°C]
dt[s]
ä
[°/s]
1:
ß
5.00
2:
ß
5.00
3:
ß
5.00
4:
ß
5.00
A ?
B header
C pgm OK
D enter
Note:
To get back to the screen with the basic parameters move the cursor to the left side of the
display labeled with “Main funct.” Alternatively you can move the cursor to the extreme right.
Temp[°C]
time
ß
#
gradient[°C]
opt
à
1: 95.0
0h 5m 0s
2: 95.0
0h 0m30s
3: 55.0
0h 0m30s
+
4: 72.0
0h 0m30s
2
A ?
B header
C pgm OK
D enter
Note:
In the main screen you will find an information that a special parameter has been
modified. This is indicated by a + in the row labeled with opt
à
.
6.1
Program time increments
To compensate for loss in enzyme activity each step within a loop can be extended from cycle to
cycle. Enter the desired time increment [seconds] in the row labeled with dt[s]. This value will
be added to the time value from cycle to cycle.
Main funct.
dT[°C]
dt[s]
ä
[°/s]
2:
ß
5.00
3:
ß
5.00
4:
ß
10
5.00
5:
ß
5.00
A ?
B header
C pgm OK
D enter