62
Biomet
®
FLX
®
Flexible Treatment Coilette
Specifications and Gauge Directions
FLX
®
Coilette Shape
Depth of Penetration
Vertical Fracture Length
Flat
2.75 cm
4 cm
Elliptical
3.5 cm
4 cm
Saddle
3.5 cm
2 cm
To insure efficacious treatment with the FLX
®
Coilette Coil:
1. Place the Coilette on the treatment site and shape for best fit.
The Coilette must only be bent in one of two directions (long axis
or short axis, see illustrations). Do not kink or twist the Coilette.
2. The shape of the Coilette will determine the coil specifications
(see above).
3. Remove the Coilette and utilize the enclosed FLX
®
Coilette Gauge
to verify the shape listed below.
Optional Shapes:
Flat Shape
(see coil specifications above)
The FLX
®
Coilette Gauge is not needed.
Elliptical Shape
(see coil specifications above)
Place the Coilette over the green portion of the
gauge marked “elliptical”. The Coilette bend
should not exceed the elliptical curve indicated
on the FLX
®
Coilette Gauge.
Saddle Shape
(see specifications above)
Place the Coilette over the red portion of the
gauge marked “saddle”. The Coilette bend
should not exceed the saddle curve on the
FLX
®
Coilette Gauge.