6 Troubleshooting
Indicator light
status
Error message
Possible cause
Action
in MANUAL mode
Action
in SLAVE mode
Flashing red
LAST SAMPLING
NOT
COMPLETED
Sampling is
interrupted in
MANUAL mode.
Press
<MENU>
to return to
the welcome message.
Press
<MENU>
to return to
the welcome message.
LAST SAMPLING
NOT
COMPLETED
Sampling is
interrupted in
SLAVE mode.
N/A
Acknowledge the fault on the
RUID and stop the sampling
cycle with the RUID that was
used to start it.
If this RUID is not
available, press
<MENU>
for
10 seconds on the instrument
and then select "
Yes
" to return
to the welcome message.
ERROR
BATTERYLIFE
There is insufficient
battery life to
perform the
sampling program.
Connect the instrument to the
mains power so that sampling
can start or put the battery on
charge.
Acknowledge the fault on the
RUID and then connect the
instrument to the mains so
that sampling can start, or to
charge the battery.
ERROR TIME >
5H
The total duration
of sequenced
samplings
programmed in
MENU 5
has
exceeded 5 hours.
Reprogram so that the total
duration does not exceed 5
hours.
Not applicable as the fault is
detected when the sampling
campaign is programmed on
the PC application.
Flashing volume
is displayed
Sampling has been
accidentally
interrupted by the
operator.
To continue sampling, press
<START>
. The volume collected
will flash on the display.
To stop sampling, press
<STOP>
.
Acknowledge the fault on the
RUID and then close the
sampling cycle with the RUID
that was used to start it.
If this RUID is not
available, press
<MENU>
for
10 seconds on the instrument
and then select "
Yes
" to return
to the welcome message.
Sampling cannot
continue. The volume collected
flashes on the instrument.
Screen is frozen
on
"MENU 6"
The battery
autonomy test
started from
MENU
6
has been
interrupted using
the
<STOP> key.
To continue the battery
autonomy test, press
<START>.
Not applicable as the battery
autonomy test can only be
performed in MANUAL mode.
REMOVE
PROTECTION
COVER
Sampling has been
interrupted as the
protective cover
has been detected.
Replace the protective cover
with the grid and then press
<MENU>
to return to the
welcome message.
If the message appears when
the protective cover has been
removed, please contact your
supplier.
Acknowledge the fault to allow
another sampling cycle to
start. Close the sampling
cycle using the RUID.
Replace the protective cover
with the sampling grid in order
to start sampling.
If the message appears when
the protective cover has been
removed, contact your