background image

 

11

 

/

 

19

 

 

I.

 

Ejection

 

Button

 

J.

 

Device

 

control

 

cap

   

On

 

the

 

lancets

 

and

 

the

 

lancing

 

device.

 

If

 

your

 

lancing

 

device

 

differs

 

from

 

the

 

one

 

shown

 

here,

 

please

 

refer

 

to

 

the

 

manufacturer’s

 

manual

 

to

 

ensure

 

proper

 

usage.

 

NOTE:

 

To

 

reduce

 

the

 

chance

 

of

 

infection:

 

● 

Never

 

share

 

a

 

lancet

 

or

 

the

 

lancing

 

device.

 

● 

Always

 

use

 

a

 

new,

 

sterile

 

lancet.

 

Lancets

 

are

 

for

 

single

 

use

 

only.

 

● 

Avoid

 

ge ng

 

hand

 

lo on,

 

oils,

 

dirt,

 

or

 

debris

 

when

 

handling

 

the

 

lancets.

 

 

Setting

 

Up

 

the

 

Lancing

 

Device

 

1.

 

Screw

 

the

 

lancing

 

device

 

cap

 

off.

 

Insert

 

a

 

needle

 

into

 

the

 

lancet

 

holder

 

and

 

push

 

down

 

firmly

 

until

 

it

 

is

 

fully

 

in

 

place.

   

 
 
 
 

2.

 

Screw

 

the

 

needle

 

cap

 

and

 

place

 

it

 

in

 

appropriate

 

location

 

to

 

use

 

again

 

after

 

finish

 

blood

taking.

 

 
 
 

3.

 

Replace

 

the

 

lancing

 

device

 

cap.

 

Turn

 

the

 

cap

 

until

 

it

 

is

 

snug

 

but

 

not

 

too

 

tight.

 

And

 

then

 

Adjust

 

the

 

adjustable

 

tip

 

to

 

proper

 

level.

 

 

NOTE:

 

The

 

adjustable

 

tip

 

offers

 

5

 

levels

 

of

 

skin

 

penetration.

 

Twist

 

the

 

adjustable

 

tip

 

to

 

the

 

proper

 

level:

 

1

2

 

level

 

for

 

soft

 

or

 

thin

 

skin

 

3

 

level

 

for

 

average

 

skin

 

4

5

 

level

 

for

 

thick

 

or

 

calloused

 

skin

 

 
 

4.

 

Pull

 

the

 

device

 

control

 

cap

 

until

 

it

 

clicks.

 

If

 

not

 

clicks,

 

check

 

the

 

needle

 

is

 

installed

 

appropriately.

     

 
 

5.

 

Aiming

 

at

 

the

 

sterilized

 

part

 

for

 

blood

taking,

 

press

 

the

 

release

 

key

 

to

 

finish

 

blood

taking.

   

 
 

6.

 

Screw

 

the

 

device

 

cap.

 

Push

 

the

 

exposed

 

needle

 

into

 

the

 

needle

 

cap,

 

and

 

then

 

push

 

the

 

ejection

 

button

 

to

 

discard

 

the

 

used

 

needle

 

in

 

Содержание G-427G

Страница 1: ...Bioland Blood Glucose Monitor Model G 427G OWNER S MANUAL...

Страница 2: ...gned to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in a...

Страница 3: ...n liquid nor put it where it could fall into liquid 2 The following basic safety notice should always be taken 3 Use the device only for the intended use described in this manual 4 Do not use adjunct...

Страница 4: ...ETER AND DELETING THE MEMORY 8 THE TWO MEASURING MODES 10 TESTING WITH BLOOD SAMPLE 10 Overview of the Lancing Device 10 Setting Up the Lancing Device 11 Preparing the puncture site 12 Performing a Bl...

Страница 5: ...e Monitoring System is intended to be used for the quantitative measurement of glucose sugar in fresh capillary whole blood samples drawn from the fingertip The Blood Glucose Monitoring System is inte...

Страница 6: ...check that the glucose meter and test strips are working properly and that you are doing the test correctly These solutions contain a known range of glucose as indicated on the strips vial bottles Thi...

Страница 7: ...cise or any other event take a blood sample from your finger only You do AST only in the following instances In a pre meal or fasting state more than 2 hours since the last meal Two hours or more afte...

Страница 8: ...and simple messages 4 M BUTTON It is used to turn on the meter enter the memory mode 5 S BUTTON It is used to set up the meter Meter Screen Display Message 1 BLOOD DROP SYMBOL Flashes when it is read...

Страница 9: ...ts might be wrong if the contact Bar is not fully inserted into the test slot The front side of the test strip should face up when inserting the test strip SETTING THE METER AND DELETING THE MEMORY 1...

Страница 10: ...n or OFF press the M button to turn on or turn off the first alarm Press the S button to select On then to set the hour When the hour is flashing press the M button to add an hour When the correct hou...

Страница 11: ...TESTING WITH BLOOD SAMPLE CAUTION To reduce the chances of infection Choose a clean dry work surface Never share a lancing device and lancet with another person Always use a new and sterile lancet La...

Страница 12: ...lace 2 Screw the needle cap and place it in appropriate location to use again after finish blood taking 3 Replace the lancing device cap Turn the cap until it is snug but not too tight And then Adjust...

Страница 13: ...l and let it air dry Fingertip testing Hold the lancing device firmly against the side of your finger Press the release key You will hear a click indicating that the puncture is complete NOTE Choose a...

Страница 14: ...hen GSM connected GSM symbol flashing and the test results data will transmission to the Cloud system automatically NOTE The results obtained from this meter are plasma calibrated This helps you and y...

Страница 15: ...Fast for at least eight hours and then do comparison tests if possible Take your meter with you to the lab While staying at the lab Make sure that the samples for both tests are taken and tested with...

Страница 16: ...the test Expired or contaminated Control Solution Expired or contaminated Test Strip Malfunc on of your Meter If you continue to get control solution test results that fall outside of the range printe...

Страница 17: ...the meter works well for example display will be clear to see after cleaning and disinfection can avoid the infection to you or to the other people and the cross infection We strongly warn that the d...

Страница 18: ...ycles Operating conditions 50 104 Humidity 25 80 RH 860hPa 1060hPa Storage Condition system meter and test strips 39 104 and below 80 relative humidity 700hPa 1060hPa Storage Condition meter 40 131 Hu...

Страница 19: ...low less than 20 or very high more than 60 red blood cell count hematocrit can lead to incorrect test results If you do not know your hemarocrit level please consult your health care provider DISPLAY...

Страница 20: ...ent or Control Solution test results are not within the specified range 1 Not enough sample in the Test Strip 2 Test Strip or Control Solution has expired 3 Test Strip has been damaged due to heat or...

Страница 21: ...20 19 Version V1 0 2016 01 by Bioland Technology LTD...

Отзывы: