4
4. When the battery is full the battery symbol disappears from the display.
If display shows RESET and PRESS TIP, press the tip ejection button to
reset the pipette, the pipette will then be ready for use.
Note:
Never reset with the tip attached as the attached tip can induce an
error in home position determination of the piston and ejector collar!
Note:
When the pipette is not in use it is recommended to keep the pipette
in the Charging Stand/Carousel. This ensures that the battery is kept charged
and the pipette is ready for use.
5.2. Charging through direct charging system
The AC-adaptor supplied with the Biohit
e
LINE Single
Charging Stand can be used for direct charging of the
Biohit
e
LINE electronic pipette. The socket for the AC-
adaptor is situated at the back of the pipette (Fig.4.).
1. Before connecting the AC-adaptor to the electrical
outlet, check that its voltage settings corresponds
to the voltage settings of your electrical outlet.
2. Check that the battery protector strip is removed
and the battery cover properly closed.
3. Connect the AC-adaptor to the electrical outlet as
well as to the socket at the back of the pipette.
4. The length of the charging time is 2.5 hours for
completely discharged batteries.
Note:
It is possible to continue pipetting while the
e
LINE pipette is charged through the AC-adaptor
(Fig. 5). However the pipette should not be operated
continuously with the AC-adaptor connected. With fully
discharged battery a 5 minute initial charging is required
before continuing the pipetting.
Fig. 5.
Fig. 4.
Fig. 3A
Fig. 3B
Fig. 3C