background image

 

 
 
 
 
 

Instructions for ThermoQ Metal Bath 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Hangzhou Bioer Technology Co., Ltd. 

Содержание ThermoQ

Страница 1: ...Instructions for ThermoQ Metal Bath Hangzhou Bioer Technology Co Ltd...

Страница 2: ...hts reserved Without prior written consent from Hangzhou Bioer Technology Co Ltd any part of the manual shall not be duplicated reproduced or translated into any other language Document No BYQ63170000...

Страница 3: ...le instrument without permission Only qualified professionals are allowed to replace elements or adjust the machine inside Replacement of elements under energizing state is prohibited d Placement of i...

Страница 4: ...ice a Warranty Content Hangzhou Bioer Technology Co Ltd is responsible for replacement for instrument with failures caused by material and manufacturing defects within 1 month after goods delivery Han...

Страница 5: ...r qualification of acceptance properly fill in the product acceptance sheet and send or fax the copied sheet to the supplier for filing and maintenance After unpacking the packing box and packing mate...

Страница 6: ...4 INSTRUCTIONS FOR INDICATOR LIGHTS AND BUTTONS 5 4 1 Instructions For Indicator Lights 5 4 2 Instructions For Buttons 5 5 OPERATION 6 5 1 Examination Before Startup 6 5 2 Start Instrument 6 5 3 Temp...

Страница 7: ...ith pressure hot lid which can prevent evaporation and explosion of lid the heating block of the hot lid is convenient for replacement Metal module avoids the sample from pollution The components of b...

Страница 8: ...ng Conditions Ambient temperature 10 C 30 C Relative humidity 70 Power supply DC24V 0 5V 90W HB 120W CHB Requirements for adapter AC100 240V 50 60HZ input current 2 5A Caution Before using the instrum...

Страница 9: ...ng Range 0 99h59min Program Segment Setting Function Yes Cycle Setting Function Yes Program Auto memory Function Yes QC Report Print Function Yes Max Segments 9 Max Cycles 99 Hot lid Operating Temp Bl...

Страница 10: ...board and functions of all buttons and preparation before startup When using the instrument for the first time read the content in the chapter carefully before startup 1 Structure Diagram Hot lid comp...

Страница 11: ...tural temperature decline status The green light is hot lid indicator light If the green light is on it shows that hot lid is under normal working status otherwise it shows that hot lid is not working...

Страница 12: ...mode PRO Multi point operation mode About 2 3 seconds later the instrument will begin to work temperature display window will show the real time temperature and corresponding indicator light will turn...

Страница 13: ...ork 2 0 Other software PDF reader Lowest configuration Processor Pentium R 4 Internal storage 1GBs Hard disc 10GB 2 Install Software Click ThermalQSetup exe in the accompanying disc to install metal b...

Страница 14: ...Instructions for ThermoQ Metal Bath Chapter IV Connect Computer 8 c Choose installation position of software click next d Choose installation folder click next...

Страница 15: ...f Complete installation click Finish 3 Online Setup 3 1 Installation of USB Drive USB drive program must be installed before online Click the accompanying disc CP210x exe install USB drive program 3...

Страница 16: ...ation The above diagram is software operation interface 1 Menu bar 2 Contents of instrument 3 Area for displaying and editing files 4 Area for displaying running status 5 Status bar 4 1 User Managemen...

Страница 17: ...tting user name click OK to save it c Delete user Choose the user to be deleted from the list of equipment click delete or choose the option of delete user with right button When deleting user delete...

Страница 18: ...e user name the system will prop up whether to override it or not when exporting it If the program document under the user is being used the user cannot be exported 4 2 File Management Include the pro...

Страница 19: ...edit or choose the option of edit file with right button After editing click save in menu bar Maximum sections 9 Maximum recycling number 99 Time scope 1min 99h59min Temperature scope 0 100 c rename...

Страница 20: ...ment click paste or the option of copy and paste in the menu of right button to paste the program document under the user name If there is the same user name with the copied one the system will prop u...

Страница 21: ...h Export file Click export choose the folder of user from the window popping up click OK to export file If there is file with the same user name the system will prop up whether to override it or not w...

Страница 22: ...display all instruments a Connect Choose sub item of equipment name in equipment list click connect or the option of instrument connection in right button menu b Disconnect Choose sub item of equipmen...

Страница 23: ...e blank and the maximum byte is 12 Instrument name can only include figure and English characters The same instrument name cannot be sued Instrument address cannot be blank and the same instrument can...

Страница 24: ...ing connected the instrument cannot be removed g Display all instruments Click all instruments the list of all instruments will appear to display the information of all instruments 4 4 Other managemen...

Страница 25: ...nt stop running local file Open the file to be ran click run the dialog box of run file will pop up Choose the instrument to be ran click OK to run the program Meanwhile the instrument will save the n...

Страница 26: ...l time temperature on upper right corner of the running status diagram it could display the real time curve C System setup Click system setup display system setup window the status of hot lid and alar...

Страница 27: ...er 21 d Check running history Click run log the form of running history will pop up e Export experimental results After the finishing of file running click to run export experimental results on upper...

Страница 28: ...y window shows 002 Fault of hot lid sensor The first time Please tune off power then tune on it Happen again Please contact supplier or manufacturer 4 Display abnormal character Fault of chip The firs...

Страница 29: ...Appendix 1 Wiring Diagram HB...

Страница 30: ...Appendix 2 Wiring Diagram CHB...

Страница 31: ...Memorandum...

Страница 32: ...ad Gaoxin Binjiang District Hangzhou P C 310053 Tel 0571 87774252 87774307 After sales Service Tel 87774558 Fax 0571 87774553 After sales Service Fax 87774559 E mail instrument bioer com cn Hot line f...