LifePro Thermal Cycler Instruction for operation
-
12
-
Press to shift between running
interface 1 and running interface 2.
1.2. How to set system parameters
Press “
System
” in main interface to access
the interface for system parameter setting.
Press
confirmation key
to move the cursor
to the next item. Roll and select the items
with “ ” by
cursor up/down key
, or press
character key
to input information for other
items, press
cursor left/right key
to move
the cursor.
Press
“
Back
” to return to main menu.
Press
“
Save
” to save parameters.
Special caution:
press “Save” to confirm
after system parameter setting or the system
parameter won’t change.
1.2.1
“Date”, “Time” indicate the default system time after system start.
1.2.2
“Default File”, “Default User” indicates the default running program name and user after
system start.
1.2.3
“Control Mode” indicates the default temperature control mode after system start, including
“Block” and “Tube” choices.
a) “Block” mode only concerns the temperature control for module during temperature
rise/fall. Now that the temperature rise/fall of reagent lags behind the module, the actual
temperature change of reagent is different from the program we set. The actual time that
the reagent reaches the set temperature is far less than the set time (see figure below for
relation between module temperature and reagent temperature).
Hotlid=off 2009-05-1 10:05:00
Date:
03-10-22
Time:
14:53:19
Default User:
PCR1
Default File:
BIOER
Control Mode:
Block
Hotlid:
OFF
Key Sound:
Yes
Run End Sound:
Yes
Back Save
Interface for system parameter setting
Hotlid=off 2009-05-1 10:05:00
100
75
50
25
0
Temp
1c 95.0
2c 55.0 3c 72.0 4
04.0
Time
1:00
02:00 01:45 00:00
File Name:PCR1
User: BIOER Now Running
•••
1c Temp=00.0 Time=00:00 Cycle1 01 of 10
Total Time: 00h 56m Remainig Time:00h 56m
Stop
Pause Start Skip View File .
module
reagent
File running interface 2