GAIT TRAINER 3 (v3.X SOFTWARE)
95
Table 11.1.
Diagnostic Error Codes Table
11.
Troubleshooting
Symptom
Gait Trainer 3 is turned ON and speed control is inoperable (belt not moving).
Check for faults on display. The following diagnostic error codes apply:
Code
Description
128
Communication between upper/lower board inop
64
Safety Lanyard Removed
32
Control fault (no amplifier)
16
Grade Error
8
Motor Tach Output Exceeds Selected Speed
4
Motor Tach Output is Below Selected Speed
NOTE:
The above faults indicate an issue with the treadmill. For fault 64-Safety Lanyard
Removed, verify lanyard is still attached in its proper location. For all other faults, please
contact Biodex Support Services department at 1-800-224-6339 and provide the fault code.