I.V. Pole (Optional Component)
The adjustable height I.V. pole should be installed onto the O.R. accessory rail. Loosen the I.V.
pole locking knob at the base of the slide block and slide the block and pole onto the accessory
rail until it is in the desired position. Tighten the locking knob to secure the pole. To adjust the
height of the I.V. pole, loosen the height adjustment locking knob and raise or lower the top
section of the pole to the appropriate height. Tighten the locking knob to secure.
Charging the Battery
This table can be operated with the power cord connected to an outlet or by using the battery.
The battery recharges automatically while the table is plugged into an outlet even if the table is
being used. An optional wall-mounted battery charger is also available.
NOTE:
The table must remain plugged into an outlet when the battery is removed or
reinstalled. If the battery is removed for any reason while the table is not plugged into a
working outlet, the battery will discharge at a faster rate. Once the battery has been
reinstalled, and it has been verified that the table is operating, the power cord plug can be
removed from the outlet and operate on battery power.
Safety Guidelines:
1.
Secure restraining straps immediately after placing the patient on the table. The
restraining straps prevent the patient from rolling off the side of the table.
NOTE:
The restraining straps are not intended to restrain an uncontrolled
patient.
2. Always lock the casters before moving a patient onto or off the table.
3. This table supports patients weighing up to a maximum of 500 pounds.