WI-108 SterilCut S28 Band Saw User Manual
Revision August 27 2019
25
Troubleshooting
Below are some common troubleshooting tips for the SterilCut S28. If these do not resolve the issue,
please contact BioCut.
Problem
Recommendations
The SterilCut S28 is not cutting material
Adjust the tensioning screw accordingly. (Pg. 12)
The SterilCut S28 is not cutting material
as desired
Adjust width of Blade Guide Bearings (Pg. 19)
The SterilCut S28 is not cutting the
material or has slowed down significantly
Clean out possible debris that has found its way into the
Drive Wheel or lower housing area.
The SterilCut S28 is stalling
Check that the blade is not over-tensioned or obstructed
by debris. (Pg. 12)
The SterilCut S28 is not running
Check to see if Blade Cover is installed correctly. (Pg. 14)
The SterilCut S28 is not running
Ensure Quick stop Switch Is not actuated (optional
feature)
The SterilCut S28 is not running
Ensure pneumatic foot pedal line(s) are intact and inserted
fully into push to connect fittings on rear of Motor Housing
The SterilCut S28 is not running
Ensure Standoff on Blade Cover is present and contacting
sensor on Motor Housing.
The SterilCut S28 is not running
Ensure Sensor is flush with Blade Housing Fixture Plate
The SterilCut S28 is not running
Unplug SterilCut S28 for at least 30 seconds then plug
back in and try again. (SterilCut S28 designed to fault in
order to protect internal components)
The SterilCut S28 is not tracking
Ensure Tensioning Wheel is not over tensioned (Pg. 12)
The SterilCut S28 is not tracking
Ensure Tensioning Wheel is not over cambered (Pg. 18)
The SterilCut S28 is not tracking
Ensure lower Socket Head Cap Screw on Carrier is
tightened
The SterilCut S28 is not tracking
If late in use of blade life, check blade for cracks, dullness
or any defects. If defect found, replace blade.
The SterilCut S28 is not tracking
Check Hub, Tension Wheel and Drive Wheel for damage
or wear
The SterilCut S28 is making abnormal
noises upon start up
Check spacing and alignment of Blade Guide and Table
Insert
PROHIBITED:
Never attempt to operate the SterilCut S28 if any part of the system has been
damaged.