Lightwave UV/Vis Diode Array Spectrophotometer
15
10.4 Cell Density/Ratio
This function key takes the user to a further screen allowing the option of cell
density or ratio measurements.
10.4.1 Cell Density Measurement
This function provides an auto-correction for use with cell cultures to
compensate for the high turbidity of such samples.
Pressing REF will set the optical density (OD) to 0.000 on a blank sample.
The cell density measurements can then be made by pressing TEST. The
results are displayed in Optical Density Absorbance.
The attenuation of the primary light beam is measured and any stray or
scattered light is rejected unlike other spectrophotometers with different
optical set ups. To compensate for this, the absorbance at two wavelengths
is measured simultaneously and an algorithm is applied providing better
comparison between instruments.