Biobase BK-HS32 Скачать руководство пользователя страница 1

Automatic Nucleic Acid Extraction System

BK-HS32

User Manual

BIOBASE GROUP

Version 2020.09

Содержание BK-HS32

Страница 1: ...Automatic Nucleic Acid Extraction System BK HS32 User Manual BIOBASE GROUP Version 2020 09...

Страница 2: ...ustry Shandong Co Ltd reserves the final right to interpret this operation manual Biobase Biodustry Shandong Co Ltd assumes responsibility for the safety and reliability of the product only if it meet...

Страница 3: ...handong Co Ltd and the operator must be systematically trained in Automatic Nucleic acid extraction system BK HS32 to meet the corresponding requirements and have the following qualifications You must...

Страница 4: ...he following professionals daily operation personnel Maintenance and troubleshooting personnel People who learn to operate The extractors instruments and systems mentioned in this article all refer to...

Страница 5: ...er precautions 10 1 3 Accessories list 10 1 4 Consumable service life 10 2 Module function introduction 11 2 1 Performance parameters 11 2 2 Module structure and function 12 2 2 1 Structure of magneti...

Страница 6: ...3 System default program 21 3 4 Program management 22 3 4 1 Language settings 23 3 4 2 Instrument Debug 23 3 4 3 Import Export 23 3 4 4 Software upgrade 23 3 5 UV disinfection 24 3 6 About 25 3 7 Exit...

Страница 7: ...ated with biological or chemical materials Heat warning label When this mark appears on the instrument it is a warning for high temperature parts be careful of hot hands Anti squeeze sign When this ma...

Страница 8: ...g the instrument without following the instructions the protective measures provided by the system may fail 1 2 1 Work environment requirements Considering the safety and performance of the instrument...

Страница 9: ...ite service This system is connected to the ground via ground wire The power ground wire must be grounded to avoid electric shock The AC power supply must be stable It is forbidden to share the power...

Страница 10: ...extend body parts into the working area of the machine otherwise it may cause injury to operators and damage to the instrument 1 2 5 Electromagnetic interference prevention To prevent electromagnetic...

Страница 11: ...ensure the stability and efficiency of the internal pollution discharge system it is necessary to replace the consumables in the instrument on time Ultraviolet lamp service life 1000 hours HEPA filter...

Страница 12: ...2 13 Heating mode The heating film 14 Temperature display resolution 0 1 15 Sample protection function Power on self check power off protection high temperature alarm overtemperature protection 16 Di...

Страница 13: ...can put the deep hole plate a total of 24 columns of hole positions of which Columns 1 8 identified in Figure 3 2 the actual hole positions are columns 1 6 7 12 13 18 19 24 are heating modules and the...

Страница 14: ...ure and normal humidity The HEPA filter can filter the aerosol generated when the instrument is heating or shaking and the dust in daily work HEPA open close method After the instrument is turned on t...

Страница 15: ......

Страница 16: ...ning The program running interface contains three function buttons Run Scan code to run and Select You can view all the programs that have been entered in the list on the left Run Click to select the...

Страница 17: ...Run interface to entering the interface as Figure 3 3 which displays the detailed steps of the program including the steps name and the time of each step Figure 3 3 3 2 2 Program running Choose a prog...

Страница 18: ...ring the test If the safety door is opened the operation will be immediately suspended until the safety door is closed and the running program can continue 2 When the program is not running the Pause...

Страница 19: ...ically resume operation until the door is manually closed 2 Under the premise that the mechanical position of the instrument has not been altered without authorization the instrument will usually not...

Страница 20: ...ure 3 7 to create a new program Click Add after Program Name to set the program name and the Add button is replaced with Cancel After entering the program name the following Save button changes from g...

Страница 21: ...the upper limit is set to 99 Blend Mode Only vertical mode is available Blend Rate Low medium and high are available Temp You can freely set whether this step needs heating and the heating temperature...

Страница 22: ...ID Step Name Description 1 Add beads The operating position is in column 2 of the deep well plate There is no waiting time during the operation The mixing time is 1 min The mixing mode is vertical os...

Страница 23: ...llation and the mixing rate is middle There is no heating demand The number of magnetization is 2 times This step The processing volume is 500ul 5 Elution The operating position is in column 6 of the...

Страница 24: ...ich have been set before leaving the factory and usually not need to be modified 3 4 3 Import Export Plug the U disk into the USB interface of the instrument to export the existing programs of the sys...

Страница 25: ...Figure 3 15 Figure 3 14 Figure 3 15 3 5 UV disinfection Click UV Operator to enter the interface shown in Figure 3 16 You can set the UV disinfection time on the left click Open to start disinfection...

Страница 26: ...rning Do not look directly at the UV lamp when it is turned on to avoid accidents 3 6 About As shown in Figure 3 17 the About interface displays the current software version machine code motor driver...

Страница 27: ...box as shown in Figure 3 19 is displayed Figure 3 19 Shut down Turn off the industrial computer Reboot Reboot the industrial computer Exit software Only exit the software the interface returns to the...

Страница 28: ...sty environment 6 During the use of the instrument please ensure that there is good ventilation around the instrument 7 Please do not use the instrument when the voltage is unstable too high or too lo...

Страница 29: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Отзывы: