Calibration Bio-Well 

Calibration instructions: 

www.bio-well.eu/forum/viewtopic.php?f=16&t=10

 

For calibration you need Calibration Pack with the following items: calibration 
stand; Ti insert; grounding cable. 

Procedure of calibration for scanning of fingers:  

1.

 

Remove Finger Insert and position Bio-Well on the calibration stand (fig.1) 

     

 

Fig.1. Bio-Well on the calibration stand. 

2.

 

Clean the optical lens with a cloth (you may use alcohol). 

3.

 

Position Ti insert and connect wires. 

4.

 

Connect Bio-Well to the computer.                           

Fig.2.

 

5.

 

In Full Scan mode test the image – it should look as a 

circle without inner noise (fig.2).  In case of inner noise, carefully adjust Ti Insert 
yellow jack to get clear image.  
6.

 

Open Calibration mode and press Start. Program will make calibration and 

make all settings (it may take up to 5-6 minutes).  
7.

 

After the process is finished, remove Ti insert.  

8.

 

You may use your Bio-Well.  

Calibration of Sputnik is being made in a different way:  

1. Instead of using white grounding cable you need to connect Sputnik antenna to 
the Ti insert. 

2. Click “Start for Sputnik” to start the calibration procedure. 