184
|
Appendix E: Instructions for Programming a New Assay
Programming the Human Anticardiolipin IgG
Enzyme Immunoassay Kit
(Standard Curve and Cutoff)
From the Main Menu, press
DEFINE
. Select the assay name and edit the name if
desired. At the define menu, follow steps below:
STEP
COMMENTS
To program the reading method, press:
METHOD
WAVELENGTH (Dual or Single): Single
MEASURE (Wavelength[s] to use): 450
PLATE TYPE: 96
To program the plate map, press:
MAP
AUTO
DOWN
DOWN
A01
BLANK MAP: AIR
NUMBER STDS: 05
NUMBER STD REPLICATES: 01
CONCENTRATIONS:
STD1: 6.25
STD2: 12.5
STD3: 25
STD4: 50
STD5:
100
NUMBER CTLS: 02
CONTROL 1: PC
CONTROL 2: NC
NUMBER OF REPLICATES
PC: 01
NC: 01
SAMPLES: 89
SAMPLE REPLICATES: 01
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...