15
9.
Empty the waste and refill the fluidics containers
using the “Hot Swap”
procedure (See Swapping Fluidics During Sorting
Final Set Up: Shut Down
The shutdown procedure is only recommended for the end of day
’s
use. If you are
not the last user, please perform a low-pressure cleaning procedure. See Final Set Up:
Low-Pressure Cleaning to do so.
1.
Fill a 5 ml tube with a minimum of 3 ml of cleaning solution (70% ethanol)
2.
Load the tube into the sample station and move it to the run position
3.
In the
Set Up and Maintenance
tab, click in the
Clean System
button
4.
A
Clean Warning
dialog will open with a prompt warning for the fact that the
tube solution will be rapidly consumed. Click in the
Yes
option
5.
In the
Clean System Modes
window, select the
High Pressure Clean
option and
click in the
checkmark
button
6.
A
Clean Warning
dialog will open with a prompt warning for the fact that the
high-pressure cleaning will leave the sorter uncalibrated. Click in the
Yes
option
7.
A
Clean System
process window will be displayed