10 | QX200 Droplet Reader and QuantaSoft Software Instruction Manual
QX200 Droplet Reader and QuantaSoft Software Instruction Manual | 11
Chapter 3 Using QuantaSoft Software
Chapter 3 Using QuantaSoft Software
3.1.1 Using the Well Editor
Use the well editor to define the settings (samples, experiment type, and detection type) for the plate. Sample
and experiment types are color-coded and can be customized for easy reference in the plate map.
1. To open the well editor, double-click on the well(s) you wish to edit. Selected wells are highlighted in gray,
and the well editor appears across the top of the interface.
■
■
To select multiple wells, hold
Ctrl
and select the wells
■
■
To select wells in a continuous series (horizontal or vertical), hold
Shift
and select the first and last wells
■
■
To select all wells in the plate, double-click in the top left corner of the plate
■
■
To select a row or column, double-click the letter or number for that row or column
2. In the Sample panel, enter the sample
Name
and select the
Experiment
from the dropdown menu.
■
■
All saved experiments appear in the dropdown menu, along with the option to
add experiment...
■
■
The sample name is case sensitive; only wells with identical sample names can be treated as merged
wells during data analysis
■
■
To create or edit an experiment, use the experiment editor (see Section 3.1.2)
3. Select the
Supermix
from the dropdown menu (required; selection cannot be changed after data collection).
Fig. 9. Assay type options.
Unused
— channel unused
Unknown
—
unknown experimental sample
Reference
— reference gene or target (required for CNV, RED, or
ratio calculations)
Positive
— positive control
Negative
— negative control
Blank
— no sample; use for wells that will not be analyzed
NTC
— no-template control
Fig. 8. Well editor.
Settings for absolute quantification of two unknowns in a single sample are shown.
Reset
— restore default settings
Apply
—
apply settings without exiting
well editor
Cancel
— close without saving changes
OK
— save changes and close well editor